Calibrateur multifonctions CX 1651
1. Calibration of industrial thermometers with Pt100 sensor and 20 mA / 10 V output:
Function setup:
2. Calibration of industrial thermometers with thermocouple sensor and frequency output:
Function setup:
3. Calibration of a single phase "electrical power/current loop" transducer
Function setup:
172
Calibrator
T RTD function
OUTPUT 140-41 ON
Multimeter
DCV or DCI depending on the output signal type of the unit
being tested
Adapter
Opt. 140-41
Calibrator
T TC function
OUTPUT 140-41 ON
Multimeter
F
Adapter
Opt. 140-41
Opt. 40 adapter
Calibrator
P function (voltage, current, power factor, frequency)
Multimeter
DCI
Adapter
Opt. 40
DUT
INPUT
OUTPUT
Pt100
H
L
DUT
INPUT
OUTPUT
TC
H
L
DUT – power transducer
INPUT
OUTPUT
HU LU
HI
LI
+I -I
Manuel de l'utilisateur