Télécharger Imprimer la page

Metrix CX 1651 Manuel De L'utilisateur page 146

Calibrateur multifonctions

Publicité

Les langues disponibles

Les langues disponibles

Calibrateur multifonctions CX 1651
No
label
error
01
Overload 2V !
02
Overload 20V !
03
Overload 200V !
04
Overload I output ! Current output overloaded
05
High temperature !
06
Overload RC !
07
FBK error !
08
OUTPUT must be
in OFF state !
10
Interface error !
11
Bad command !
12
Bad
communication !
13
Over range !
20
Bad calib. code !
21
Time warm up !
24
Cable adapter must
be off !
25
Use cable adapter ! Attempt to start calibration
30
Internal RxD
timeout !
31
Internal
communication !
37
Calibrator is not
ready !
40
Value too large !
41
Value too small !
42
Deviation too large
!
44
Unable +/- !
45
Unable – polarity !
46
Unable DC/AC !
47
Current timeout
!
48
Not allowed on
AUX output !
146
description
2V range overloaded
20 V range overloaded
200, 1000 V ranges
overloaded
Too high internal
temperature
RC simulator overloaded
Internal error
Cable adapter tried to be
exchanged while output
terminals was ON
GPIB communication error Wrong data format on GPIB.
Bad command of GPIB
GPIB communication error Listener not connected to the GPIB. Check correct connection of
Overcrossing the range via
GPIB
Bad calibration code
Attempt to start calibration
before warm up
Adapter isn't allowed for
autocalibration
without cable adapter.
Internal error
Internal error
Internal error
Maximum value is out of
limit
Minimum value is out of
limit
Deviation is too high
Change of polarity is not
allowed
Negative polarity is not
allowed.
DC/AC conversion is not
possible
Time limit for current over
10 A exceeded
Function cannot be used
on output AUXILIARY
troubleshooting
Output current is too high. Increase load resistance.
Output current is too high. Increase load resistance.
Output current is too high. Increase load resistance.
Voltage on the load is too high. Decrease load resistance.
Output stages are overloaded. Do not use ranges 200V, 1000V or
20 A for at least 10 minutes. Check if the ventilation holes are
free.
Test current is too high. Use lower range of tested Ohmmeter.
Turn off the calibrator and turn on it again.
Switch off output terminals with button OUTPUT, change the
adapter, and switch output terminals on.
Not known command on GPIB.
GPIB cable.
Value out of range was set via GPIB. Set correct value.
Wrong calibration code was entered, calibration cannot start.
Enter correct calibration code.
Attempt to start calibration before 60 minutes warm up period.
Let the calibrator turned on for at least 60 minutes.
Use another cable adapter or perform autocalibration procedure
without cable adapter.
Calibration of resistance ranges can be performed with adapter
Option 70. Calibration of internal multimeter can be performed
with cable adapters Option 40 and Option 60.
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
Internal error of the calibrator. Turn the calibrator off and after 5
s turn on. If the error will appear again, contact manufacturer.
Attempt to set value over possible range. Set correct value.
Attempt to set value under possible range. Set correct value.
Set deviation is out of limit –30% to +30%. Set correct value.
Attempt to change polarity, where it is not allowed.. Concerned
modes F, P-E, R-C, ACV, ACI.
Attempt to set negative polarity, where it is not allowed..
Concerned modes F, P-E, R-C, ACV, ACI.
Attempt to change parameter AC/DC where it is nonsense or
where it is not allowed.
Long-term loading current terminals with output current over A.
Do not use this setting in combination with connected cable
adapter.
Manuel de l'utilisateur

Publicité

Chapitres

loading

Produits Connexes pour Metrix CX 1651