Manufacturer specification limits
Finnpipette Digital Single Channel
Range
Volume
µl
0.2-2 µl
2
0,2
0.5-10 µl, micro 10
1
0.5-10 µl
10
1
2-20 µl
20
2
5-50 µl
50
5
10-100 µl
100
10
20-200 µl
200
20
100-1000 µl
1000
100
1-5 ml
5000
1000
2-10 ml
10000
2000
Finnpipette Digital Multichannel
Range
Channels
0.5-10 µl
8, 12, 16
5-50 µl
8, 12, 16
50-300 µl
8, 12
ADJUSTMENT
Adjustment is done only for one volume. The recommended
adjustment volume is the minimum volume or 10% of the maximum
volume. Multichannel pipettes adjustment check prior calibration
is done using one of the middle channels.
1. Place the service tool into the openings of the calibration nut at
the top of the handle.
2. Turn the service tool clockwise to increase, or counterclockwise
to decrease the volume.
3. After adjustment check the calibration according to the
instructions above.
FORMULAS FOR CALCULATING RESULTS
Conversion of mass to volume
V = (w + e) x Z
Inaccuracy
µl
%
±0,050
±2,50
±0,024
±12,00
±0,100
±1,00
±0,025
±2,50
±0,100
±1,00
±0,035
±3,50
±0,20
±1,00
±0,06
±3,00
±0,30
±0,60
±0,15
±3,00
±0,80
±0,80
±0,30
±3,00
±1,2
±0,60
±0,36
±1,80
±6,0
±0,60
±1,0
±1,00
±25,0
±0,50
±8,0
±0,80
±50,0
±0,50
±20,0
±1,00
Volume
Inaccuracy
µl
µl
10
±0,240
1
±0,120
50
±0,75
5
±0,25
300
±3,0
50
±2,3
V = volume (µl)
w = weight (mg)
e = evaporation loss (mg)
Z = conversion factor for mg/µl conversion
Imprecision
s.d. µl
cv% calibration tip
0,040
2,00
0,020
10,00
0,050
0,50
0,020
2,00
0,080
0,80
0,030
3,00
0,08
0,40
0,05
2,50
0,15
0,30
0,125
2,50
0,20
0,20
0,10
1,00
0,4
0,20
0,14
0,70
2,0
0,20
0,6
0,60
10,0
0,20
3,0
0,30
20,0
0,20
6,0
0,30
Imprecision Factory
%
s.d. µl
cv% calibration tip
±2,40
0,160
1,60
±12,00
0,080
8,00
±1,50
0,35
0,70
±5,00
0,10
2,00
±1,00
0,9
0,30
±4,60
0,75
1,50
9
Factory
10
10
250 Univ.
250 Univ.
250 Univ.
250 Univ.
250 Univ.
1000
5 ml
10 ml
10
250 Univ.
300