Technical data
Changes reserved!
Performance specifications
Operating mode
Transmittance (%), Absorbance and Concentration
Deuterium lamp (UV) and halogen lamp (visual
Source lamp
range)
Wavelength range
190–1100 nm
Wavelength Accuracy
± 1 nm (wavelength range 200–900 nm)
Wavelength
< 0.1 nm
reproducibility
Wavelength resolution
0.1 nm
Wavelength calibration
Automatic
Wavelength selection
Automatic, based on method selection
Scanning speed
900 nm/min (in 1-nm steps)
Spectral bandwidth
2 nm (1.5–2.9 nm at 656 nm, 1 nm for D2 line)
Photometric measuring
± 3 Abs (wavelength range 200–900 nm)
range
5 mAbs at 0.0–0.5 Abs
Photometric accuracy
< 1% at 0.50–2.0 Abs at 546 nm
< 0.5% to 2 Abs
Photometric linearity
≤ 1% at > 2 Abs with neutral glass at 546 nm
KI solution at 220 nm
Stray light
< 3.3 Abs / < 0.05%
190–199 nm +/- 0.0100 Abs
Photometric drift against
200–349 nm +/- 0.0054 Abs
blank reading
350–899 nm +/- 0.0034 Abs
(30-min baseline stable)
900–1100 nm +/- 0.0100 Abs
Performance specifications
Long-term stability
Zero point at 546 nm for 10 hours ≤ 0.0034 Abs
5000 measurement values (result, date, time, sample
Data log
ID, user ID)
50 scans, 50 time scans
User programs
200
Physical and environmental specifications
Width
500 mm (19.69 in)
Height
215 mm (8.46 in)
Depth
460 mm (18.11 in)
Weight
11 kg (24.25 lb)
10–40 °C (50–104 °F), maximum 80% relative
Ambient operating
humidity
requirements
(without condensate formation)
–25–60 °C (–13–140 °F), maximum 80% relative
Ambient storage
humidity
requirements
(without condensate formation)
Additional technical data
Power connection
100–240 V/50–60 Hz
Power consumption
150 VA
Fuse
T 2A H; 250 V (2 units)
Use only shielded cable with maximum length of 3 m:
2x USB type A
1x USB type B
Interfaces
Use only shielded cable (for example STP, FTP, S/
FTP)
with maximum length of 20 m:
1x Ethernet
Housing rating
IP20 with closed cell compartment cover
English
English 3