6. Operation

6.8. Displaying calibration data

To view the data obtained after
calibration on screen.
Note:
- If the pH was calibrated at three
points, the data for the two slopes
obtained will be displayed.
- During conductivity calibration,
the instrument adjusts a constant
for each standard. Hence, a single
constant will be displayed if one-
point calibration was performed, 2
constants in the case of two-point
calibration and 3 constants for three-
point calibration.
(1) Press
to select the required option.
(2) Press the
neously.
sensION+ MM150 HACH 0710.indd ANG:19
sensION+ MM150 HACH 0710.indd ANG:19
and
(1)
keys simulta-
(2)
Slope
Asymmetry potential
Sensitivity (%)
Calibration temperature
pH
Conductivity (EC)
(1)
(2)
Constant and
temperature
1st standard
etc.
REDOX (ORP)
(1)
(2)
Deviation
obtained
Calibration
temperature
etc.
19