Lamp Replacement - Hach DR 6000 Manuel D'utilisation De Base

Masquer les pouces Voir aussi pour DR 6000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 59
Table 6 Time course options (Continued)
Options
Description
Scale: In the automatic Scaling mode, the y-axis is
automatically adjusted so that the total scan is displayed.
Scale & Units
The manual Scaling mode allows sections of the scan to be
displayed.
Units: Choice of absorbance or transmittance.
To send Data to a printer, computer or USB memory stick
Send Data
(Type A)
Single Reading: One measurement result is shown after
tapping on Read.
Carousel 1 inch square: Optional measurement of carousel
Reading Mode
insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.
Recall saved measurement data, wave-length scans or time
Recall
courses, see
Save, recall, send and delete data, page
measurement
data
Instrument Setup Basic data of the instrument.
English 26
Maintenance
Potential Chemical, Biological Eye and Skin Hazards.
Only qualified personnel should conduct the tasks described in this section of the
manual.
Remove all cells remaining in the instrument. Dispose of the cells or their
contents in an authorised manner.

Lamp replacement

Risk of electric shock
Disconnect the instrument from the power source before the lamp replacement
procedure is commenced and until the lamp replacement procedure is complete.
Burn hazard.
Wait until the lamp cools down. Contact with the hot lamp can cause burns.
14.
Only hold lamp at lamp holder. Avoid touching the glass, as substances on the
skin can bake onto the lamp bulb and thus accelerate the ageing process of the
lamp.
C A U T I O N
N O T I C E
D A N G E R
C A U T I O N
N O T I C E

Hide quick links:

Publicité

Table des Matières
loading

Table des Matières