3.8 Sensor status

The Sensor status screen is shown only when the gap between the inductive proximity
sensors and the reference bar is not within the recommended range of 3 ±0.5 mm
(0.12 in ±0.02 in) or when the battery level of the measuring unit is below 10%.
a.
Connected and serial number
Indicate if a TKSA 11 measuring unit is connected and shows
its serial number.
b.
Battery level
Indicates the charge level for the internal battery. A warning is
displayed when below 10%.
c.
Probe distances (mm or mils)
In order to obtain the largest possible range, it is
advisable to start the measurement at 3mm ±0.5 mm
(120 mils ±20 mils) distance between the sensors and
reference bar. Values displayed in red are not within the
recommended range. The two arrows show how to adjust the
measuring unit in the correct direction.
Note:
The yellow warning sign indicates at least one of the distances
is not within the advised range.
Note:
The red "Stop" sign indicates the sensors are too far or too
close to the reference bar to get a reading.
Setup assistance
Indicates what caused the warning or stop, and gives clues on how to ix the problem.
d.
Rotation angle
The rotational angle for the unit is displayed and can be used for precise positioning.