26. TECHNICAL DATA

Display .................................Max. 60000 counts (characters)
Measuring rate ..................... approx. 2 – 3 measurements/second,
Measuring line length ..........approx. 90 cm each
Measuring impedance .........>10 MΩ (V range)
Operating voltage ................9 V block battery
Working conditions ..............0 to +30 ºC (<75 % RH), +30 to +40 ºC (<50 % RH)
Operating altitude ................max. 2000 m
Storage temperature ............-10 to +50 ºC
Weight .................................approx. 380 g
Dimensions (L x W x H) .......185 x 91 x 43 mm
Over-voltage category .........CAT III 1000 V, CAT IV 600 V, contamination degree 2
Measurement tolerances
Statement of accuracy in ± (% of reading + display error in counts (= number of smallest points)). The
accuracy is valid for one year at a temperature of +23 °C (± 5 °C), and at a relative humidity of less
than 75 %, non-condensing. Temperature coefficient: +0.1 x (specified accuracy)/1 °C
Direct voltage
Range
600 mV
6 V
60 V
600 V
1000 V
Overload protection: 1000 V; Impedance: 10 MΩ
Bargraph: approx. 2 – 3 measurements/second
Accuracy
±(0,03 % + 10 digits)
±(0,05 % + 10 digits)
Resolution
0.01 mV
0.0001 V
0.001 V
0.01 V
0.1 V
81