Covidien Edge E1455 Mode D'emploi page 5

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
Warning
Some surgeons may elect to "buzz the
hemostat" during surgical procedures. It is
not recommended, and the hazards of such
a practice probably cannot be eliminated.
Burns to the surgeon's hands may result. To
minimize the risk, take these precautions:
• Do not "buzz the hemostat" with a needle
electrode.
• Do not lean on the patient, the table, or
the retractors while buzzing the hemostat.
• Activate cut rather than coag. Cut has
a lower voltage than coag.
• Use the lowest power setting possible for
the minimum time necessary to achieve
hemostasis.
• Activate the generator after the accessory
makes contact with the hemostat. Do not
arc to the hemostat.
• Firmly grasp as much of the hemostat as
possible before activating the generator.
This disperses the current over a larger
area and minimizes the current
concentration at the fingertips.
Warning
• "Buzz the hemostat" below hand level (as
close as possible to the patient) to reduce
the opportunity for current to follow
alternate paths through the surgeon's
hands.
• When using a coated or nonstick blade
electrode, place the edge of the electrode
against the hemostat or other metal
instrument.
Precaution
Do not exceed maximum power limits as
stated in instructions for use. Exceeding
these power settings may result in patient
injury or product damage.
Always use the lowest power setting that
achieves the desired surgical effect. Use
the active electrode for the minimum time
necessary in order to reduce the possibility
of unintended burn injury.
en
5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Edge e1455-4Edge e1455-6Edge e1465Edge e1465-4Edge e1465-6

Table des Matières