Order: 6334
Type
version
6334
Standard
ATEX Ex
Electrical specifications
Specifications range:
-40°C to +60°C
Common specifications:
Supply voltage, DC
Standard............................................... 7.2...35 V
ATEX Ex ................................................ 7.2...30 VDC
Internal consumption .................................. 0.17...0.8 W
Voltage drop ................................................ 7.2 VDC
Isolation voltage, test / operation ............... 1.5 kVAC / 50 VAC
Isolation voltage, channel 1 / channel 2:
Standard............................................... 3.75 kVAC
ATEX Ex ................................................ 1500 VAC
Warm-up time .............................................. 5 min.
Communications interface .......................... Loop Link
Signal / noise ratio ...................................... Min. 60 dB
Response time (programmable) .................. 1...60 s
EEprom error check .................................... < 3.5 s
Signal dynamics, input ................................ 18 bit
Signal dynamics, output.............................. 16 bit
Calibration temperature............................... 20...28°C
Accuracy, the greater of general and basic values:
Input type
All
All manuals and user guides at all-guides.com
: A
1500 VAC
: B
General values
Absolute
accuracy
≤ ±0.05% of span
6334V105
Galvanic
isolation
: 2
Single
Double
Temperature
coefficient
≤ ±0.01% of span / °C
Channels
: A
: B
17