Sommaire des Matières pour Huntleigh Smartsigns LITE Serie
Page 1
All manuals and user guides at all-guides.com HEALTHCARE SL500 / SL510 HEALTHCARE START CYCLE STOP REVIEW CLEAR S S m m a a r r t t s s i i g g n n s s ® ® L L I I T T E E S S e e r r i i e e s s USER MANUAL...
Page 2
All manuals and user guides at all-guides.com English ..3 - 35 Français ..36 - 45 Deutsch ..46 - 55 Español .
Page 3
All manuals and user guides at all-guides.com HEALTHCARE LITE English Quick Reference Guide & User Manual Contents Page No. Quick Reference Guide ....5 Quality, Reliability and Safety .
Page 4
All manuals and user guides at all-guides.com HEALTHCARE Contents Page No. Thermometry Operating Instructions ... .18 Select Appropriate Temperature Probe ....18 Installing Temperature Probe Cover .
Page 5
All manuals and user guides at all-guides.com HEALTHCARE Quick Reference Guide This guide is intended as a quick reference for experienced users. All users should review and understand the full operating instructions as detailed in this User Manual. Power ON/OFF Locate the black power switch on the back surface near the top right (as viewed from the front of the device).
Page 6
All manuals and user guides at all-guides.com HEALTHCARE Error Codes M M o o d d e e C C o o d d e e D D e e s s c c r r i i p p t t i i o o n n C C o o r r r r e e c c t t i i v v e e A A c c t t i i o o n n - - - User-cancelled...
Page 7
All manuals and user guides at all-guides.com HEALTHCARE User Manual 1. Quality, Reliability and Safety T T h h i i s s U U s s e e r r M M a a n n u u a a l l i i f f d d e e s s i i g g n n e e d d f f o o r r u u s s e e w w i i t t h h t t h h e e S S m m a a r r t t s s i i g g n n s s ® ® L L I I T T E E S S e e r r i i e e s s V V i i t t a a l l S S i i g g n n s s M M o o n n i i t t o o r r s s - S S L L 5 5 0 0 0 0 a a n n d d S S L L 5 5 1 1 0 0 .
Page 8
All manuals and user guides at all-guides.com HEALTHCARE D D o o n n o o t t substitute accessories. Use only recommended P P o o s s s s i i b b l l e e accessories listed in this manual. Substitution may cause S S a a f f e e t t y y R R i i s s k k the instrument to work improperly.
Page 9
Smartsigns® LITE Series. Preliminary Checks Huntleigh Healthcare takes every precaution to ensure that their goods reach you in perfect condition. We recommend that a thorough visual inspection is made prior to installation. Should any damage be evident or any parts missing, ensure that Huntleigh Healthcare, Diagnostic Products Division are informed at once.
Page 10
All manuals and user guides at all-guides.com HEALTHCARE 2. Description of the Devices The Smartsigns® LITE Series are a range of automated, non-invasive blood pressure monitors, designed to measure blood pressure using an upper arm cuff. Recommended Clinical Applications The Smartsigns® LITE Series measures systolic and diastolic blood pressures and pulse rates in subjects three years of age or older.
Page 11
All manuals and user guides at all-guides.com HEALTHCARE Display Descriptions The following displays are available on the SL500 and SL510 models: S S Y Y S S T T O O L L I I C C pressure in millimetres of mercury (mmHg). D D I I A A S S T T O O L L I I C C pressure in millimetres of mercury (mmHg).
Page 12
All manuals and user guides at all-guides.com HEALTHCARE Smartsigns® LITE Button Descriptions The following buttons are available on the SL500 and SL510 models: NIBP Button Description START Starts blood pressure measurement(s). STOP Suspends Automatic Mode. (Press START to resume.) Clears memory of all stored readings in manual and CLEAR automatic mode.
Page 13
All manuals and user guides at all-guides.com HEALTHCARE 3. Setup Instructions Accessories Each system is supplied with the following items:- Smartsigns® LITE Series Automated Power supply Non-Invasive Blood Pressure Monitor Coiled tubing Mounting bracket and hardware Three blood pressure cuffs Instrument warranty card User manual Temperature Option:...
Page 14
All manuals and user guides at all-guides.com HEALTHCARE Attaching the Temperature Probe to the SL510 Connecting the cuff Push the hose onto the connector on the front of the unit (Fig.1). Connect the hose to the cuff via the threaded START connector.
Page 15
All manuals and user guides at all-guides.com HEALTHCARE Power-Up 1. Switch On 2. LED will illuminate. 4. Unit is ready 3. Tone will sound START CYCLE START CYCLE STOP REVIEW STOP REVIEW CLEAR CLEAR NOTE If the unit is left idle for 10 minutes, all displays will go blank. Press any button to re-enter the normal operating mode.
Page 16
R R a a t t e e r r a a n n g g e e a a n n d d a a c c c c u u r r a a c c y y i i n n f f o o r r m m a a t t i i o o n n . . Select Appropriate Cuff Size Selecting the appropriate cuff size is important, otherwise blood pressure readings may be inaccurate. Huntleigh Healthcare Ltd. provide five cuff sizes. Child 13 - 18 cm Small adult...
Page 17
All manuals and user guides at all-guides.com HEALTHCARE Automatic Mode - Clear - until required cycle time is displayed (1-5 minutes). - Start Cuff Cuff Inflates Pressure Displayed START CYCLE STOP REVIEW CLEAR Cuff Measurement Deflates Displayed START CYCLE STOP REVIEW CLEAR Average...
Page 18
All manuals and user guides at all-guides.com HEALTHCARE Thermometry Operating Instructions (SL510) T T o o g g g g l l e e b b e e t t w w e e e e n n C C e e l l s s i i u u s s a a n n d d F F a a h h r r e e n n h h e e i i t t b b y y p p r r e e s s s s i i n n g g a a n n d d h h o o l l d d i i n n g g t t h h e e M M O O D D E E k k e e y y d d u u r r i i n n g g P P o o w w e e r r U U p p .
Page 19
All manuals and user guides at all-guides.com HEALTHCARE -select - start measurement. The progress Single Single indicators will oral axillary alternate O O r r a a l l A A x x i i l l l l a a r r y y After approx.
Page 20
All manuals and user guides at all-guides.com HEALTHCARE 6. Technical Information General Specification Item Pressure measurement range 0 to 290 mmHg Meets or exceeds ANSI/AAMI SP10:1992 Pressure measurement Within 3 mmHg or 2%, whichever is greater accuracy Systolic blood pressure 60 to 250 mmHg measurement range Diastolic blood pressure...
Page 21
All manuals and user guides at all-guides.com HEALTHCARE SL500-BAT & SL510 Only Item Specification Input power requirements to 100 - 240 VAC, 390 mA, 50 - 60 Hz universal power supply Output power of universal 10 V, 1.25 A power supply Cordset connector type IEC 320 C7 connector 2 m, 6.7 feet...
Page 22
Smartsigns® LITE . The users' application can command the Smartsigns® LITE to perform an NIBP measurement or a temperature measurement and collect data into the user's PC-based application. Contact Huntleigh Healthcare to obtain the Smartsigns® LITE connectivity manual.
Page 23
All manuals and user guides at all-guides.com HEALTHCARE 7. Cleaning, Maintenance and Storage Care of your Equipment Although the Smartsigns® LITE Series is robust and has been designed to withstand normal clinical use, the unit contains delicate components which should be handled and treated with care. We recommend that the system is included into an annual calibration programme where the accuracy of the system is checked against the manufacturers' specifications.
Page 24
All manuals and user guides at all-guides.com HEALTHCARE Maintenance Huntleigh Healthcare recommends that preventative maintenance checks are carried out at least annually. Maintenance must only be carried out by suitably qualified personnel. Replacing the Battery (SL500-BAT/SL510) B B a a t t t t e e r r y y C C o o v v e e r r...
Page 25
HEALTHCARE 8. Warranty and Service Information Warranty Huntleigh Healthcare's standard terms and conditions apply to all sales. A copy is available on request. These contain full details of warranty terms and do not limit the statutory rights of the consumer.
Page 26
All manuals and user guides at all-guides.com HEALTHCARE Error Messages and Trouble Shooting NIBP System Errors If a NIBP system error occurs during operation, an error code will be displayed in the SYSTOLIC display only. If an NIBP system error (indicated by '*' in the table below) occurs during automatic mode, the error code is stored in memory and the series automatically continues until six measurements are performed.
Page 27
All manuals and user guides at all-guides.com HEALTHCARE NIBP System Description Corrective Action Error Code Readings in memory, if any, are available E 13 Device reset. for review. Check device environment for electro-magnetic interference (ESD, power interruptions). Connect USB cable between measurements or when device is powered off.
Page 28
All manuals and user guides at all-guides.com HEALTHCARE 9.3 Temperature Errors Thermometry Corrective Action Description Error Code Continuous temperature E 41 Re-calibrate over display limit (>43°C) Continuous temperature E 42 Re-calibrate under display limit (<28°C) No temperature probe Attach temperature probe and repeat E 43 connected measurement.
Page 29
All manuals and user guides at all-guides.com HEALTHCARE Unrecoverable System Errors Safety mechanisms in the Smartsigns® LITE Series are designed to monitor for particular unrecoverable system errors in which the device may fail to operate normally and may produce a possible hazard to the user or patient. If an unrecoverable system error occurs, the device will:- Deflate the cuff.
Page 30
All manuals and user guides at all-guides.com HEALTHCARE Troubleshooting Guide Troubleshooting Corrective Action Possible Cause(s) Symptoms Device is in Automatic Mode Press CLEAR to clear the Power is on (green and 6 readings have already memory, and then press indicator light is been taken.
Page 31
Pulse and Reading displays show varying numbers. CAUTION: If the troubleshooting information in this manual does not solve the problem, contact Huntleigh Healthcare Ltd. or your distributor for technical service or replacement of the Smartsigns® LITE Series or accessories.
Page 32
All manuals and user guides at all-guides.com HEALTHCARE 10. Accessories WARNING: The use of a power supply other than that supplied with the Smartsigns® LITE Series may result in damage to the device and/or measurement inaccuracy. WARNING: Substitution of a component, such as a blood pressure cuff, temperature probe or probe cover different from that supplied may result in measurement inaccuracy.
Page 33
Align unit. Push towards the bracket and down to lock. Manufactured By: Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with:...
Page 34
All manuals and user guides at all-guides.com HEALTHCARE SL500-B B AT T / SL510 S S U U P P P P L L I I E E D D F F I I X X I I N N G G S S 4 4 x x 4 4 x x 4 4 x x...
Page 35
All manuals and user guides at all-guides.com HEALTHCARE 11.2 Mounting to the Mobile Stand (SL500-BAT & SL510) S S U U P P P P L L I I E E D D F F I I X X I I N N G G S S 1 1 x x 6 6 x x 4 4 x x...
Page 36
All manuals and user guides at all-guides.com HEALTHCARE Français Guide de référence rapide Table des matières Page 1. Avertissements ......36 2.
Page 37
All manuals and user guides at all-guides.com HEALTHCARE Ne jamais immerger une portion de l'instrument dans de E E n n d d o o m m a a g g e e m l'eau. Un contact avec des liquides peut endommager les e e n n t t p p o o s s s s i i b b l l e e d d e e composantes électriques de l'instrument.
Page 38
All manuals and user guides at all-guides.com HEALTHCARE Installation Contenu Prise secteur N.B. Chargez la batterie ... pendant au moins 12 heures avant la première utilisation. pendant toute la nuit si l'unité est utilisée en mode batterie pendant plus de 2 heures. au moins une fois tous les trois mois lorsque l'unité...
Page 39
All manuals and user guides at all-guides.com HEALTHCARE Connexion du brassard Insérez le flexible dans le connecteur START Connecteur du brassard Fixation de la thermosonde (SL510) Mise sous tension Allumez Le DEL s'allume Un son se fera entendre L'unité est prête START CYCLE REVIEW...
Page 40
All manuals and user guides at all-guides.com HEALTHCARE Fonctionnement Prise de mesure de NIBP P P l l a a c c e e z z l l e e b b r r a a s s s s a a r r d d s s u u r r l l e e p p a a t t i i e e n n t t . . N N .
Page 41
All manuals and user guides at all-guides.com HEALTHCARE Les étapes 4 - 5 sont répétées 6 6 fois. Un son se fera entendre. La mesure moyenne est affichée. START CYCLE Visualisation des lectures précédentes - pour afficher la lecture précédente (le numéro dans l'affichage "READING "...
Page 42
All manuals and user guides at all-guides.com HEALTHCARE Mode simple N.B.: Pour une prise de mesure simple, appuyez sur avant d'appliquer la sonde à la partie du corps choisie. Assurez-vous que la portion métallique de la sonde n'est pas en contact avec la peau avant la prise de mesure. - Sélectionnez - Démarrage de la prise de mesure...
Page 43
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 44
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Alignez l'unité. Poussez en direction du support, puis HEALTHCARE vers le bas pour la fixer. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Montage sur le support mobile (SL500-BAT &...
Page 45
été nettoyé) à l'extérieur de l'emballage. Écrire sur l'emballage : " Service Department - Smartsigns® LITE ". La société Huntleigh Diagnostics se réserve le droit de renvoyer tout produit ne présentant pas un certificat de décontamination. Un manuel d'entretien est disponible pour ce système et contient des informations d'entretien,listes des pièces et conseils pour le diagnostic.
Page 46
All manuals and user guides at all-guides.com HEALTHCARE Deutsch Kurzes Handbuch Inhalt Seitenzahl 1. Warnhinweise ......46 2.
Page 47
All manuals and user guides at all-guides.com HEALTHCARE Tauchen Sie nie Teile oder das gesamte Gerät in Wasser M M ö ö g g l l i i c c h h e e ein. Verschüttete Flüssigkeiten können die elektrischen G G e e r r ä...
Page 48
All manuals and user guides at all-guides.com HEALTHCARE Installation Inhalt Netzanschluss ANMERKUNG Laden Sie die Batterie ... mindestens 12 Stunden vor Erstgebrauch auf. übernacht auf, wenn das Gerät mehr als 2 Stunden lang Batterie betrieben werden soll. mindestens ein Mal alle drei Monate auf, um die Betriebsdauer der Batterie zu verlängern.
Page 49
All manuals and user guides at all-guides.com HEALTHCARE Manschette anschließen Stecken Sie den Schlauch auf das START Verbindungs-stück Verbindungsstück Temperatursonde anschließen (SL510) Einschalten Anschalten LED leuchtet auf Ton wird hörbar Einheit betriebsbereit START CYCLE REVIEW STOP START CYCLE CLEAR STOP REVIEW CLEAR ANMERKUNG...
Page 50
All manuals and user guides at all-guides.com HEALTHCARE Betrieb NIBP-Messung D D e e m m P P a a t t i i e e n n t t e e n n M M a a n n s s c c h h e e t t t t e e a a n n l l e e g g e e n n . . A A n n m m e e r r k k u u n n g g : : s s t t o o p p p p t t d d a a s s M M e e s s s s e e n n u u n n d d l l ä...
Page 51
All manuals and user guides at all-guides.com HEALTHCARE Wird 4 - 5 6 6 Mal wiederholt. Ton erklingt. Anzeige durchschnittlicher Wert. Vorherige Ergebnisse einsehen START CYCLE - zeigt vorheriges Ergebnis (die Zahl auf dem ERGEBNIS [READING] stimmt mit dem angezeigten Messergebnis überein). - löscht gespeicherte Ergebnisse.
Page 52
All manuals and user guides at all-guides.com HEALTHCARE Einzelbetrieb ANMERKUNG: Drücken Sie bei Einzelmessungen auf , bevor Sie die Sonde am Messort anbringen. Gewährleisten Sie, dass das Metallteil der Sonde vor der Messung nicht in Kontakt mit der Haut kommt. - Messung - Auswählen beginnen...
Page 53
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 54
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Einheit ausrichten. In Richtung Konsole und nach unten zum HEALTHCARE Einrasten drücken. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Auf beweglichem Ständer montieren (SL500-BAT &...
Page 55
All manuals and user guides at all-guides.com HEALTHCARE Gewährleistung und Kundendienst Für alle Verkäufe gelten die allgemeinen Verkaufsbedingungen von Huntleigh Healthcare Diagnostic Products Divisions. Exemplare auf Anfrage erhältlich. Hierin sind die genauen Einzelheiten der Gewährleistungsbedingungen angegeben und diese beschränken nicht die gesetzlichen Rechte des Verbrauchers.
Page 56
All manuals and user guides at all-guides.com HEALTHCARE Español Guía de Referencia Rápida Contenidos Nº de Página 1. Advertencias ......56 2.
Page 57
All manuals and user guides at all-guides.com HEALTHCARE P P o o s s i i b b l l e e D D a a ñ ñ o o d d e e l l N N o o introduzca ningún componente del aparto en agua. El E E q q u u i i p p o o líquido al extenderse puede dañar los componentes eléctricos del aparato.
Page 58
All manuals and user guides at all-guides.com HEALTHCARE Configuración Índice Conexión a la red eléctrica NOTA Cargue la batería… ... al menos 12 horas antes del uso inicial..por la noche si el aparato se usa en modo batería durante más de 2 horas..
Page 59
All manuals and user guides at all-guides.com HEALTHCARE Conexión del manguito Presione el tubo START contra el conector Manguito del conector Conexión de la Sonda de Temperatura (SL510) Encendido Pulse Interruptor Se encenderá el Emitirá un tono audible Unidad preparada START CYCLE STOP...
Page 60
All manuals and user guides at all-guides.com HEALTHCARE Funcionamiento Medida de la TA no invasiva C C o o l l o o q q u u e e e e l l m m a a n n g g u u i i t t o o a a l l p p a a c c i i e e n n t t e e . . N N o o t t a a : : d d e e t t i i e e n n e e l l a a m m e e d d i i c c i i ó...
Page 61
All manuals and user guides at all-guides.com HEALTHCARE 4 - 5 repetidos 6 veces. Se emitirá un tono audible. Se visualiza la medición media. START CYCLE Revisión de lecturas anteriores - muestra la lectura anterior, (el número de pantalla de LECTURA corresponderá...
Page 62
All manuals and user guides at all-guides.com HEALTHCARE Modo Aislado NOTA: Para medidas aisladas, pulse antes de colocar la sonda en el lugar de la medición. Antes de realizar la medición, compruebe que la parte metálica de la sonda no está en contacto con la piel prior. -seleccionar - iniciar la medición...
Page 63
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 64
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Unidad de alineación. Presione hacia el soporte y HEALTHCARE hacia abajo hasta encajarlo. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Montaje sobre el Soporte Móvil (SL500-BAT &...
Page 65
Anote en el exterior del embalaje "Servicio Técnico - Smartsigns LITE". Huntleigh Diagnostics se reserva el derecho a devolver cualquier producto que no incluya un certificado de descontaminación. Ponemos a su disposición un manual técnico para el equipo que incluye información técnica, listados de los componentes y consejos para detectar...
Page 66
Q Q ualsiasi modifica o riparazione deve essere effettuata da un tecnico qualificato o da un agente o un tecnico ospedaliero autorizzato da Huntleigh Healthcare. 1. Avvertimenti Q Q u u a a n n t t o o s s e e g g u u e e è è l l a a d d e e s s c c r r i i z z i i o o n n e e d d i i r r i i s s c c h h i i g g e e n n e e r r a a l l i i e e d d i i...
Page 67
All manuals and user guides at all-guides.com HEALTHCARE Non immergere alcuna parte dello strumento in acqua. P P o o s s s s i i b b i i l l e e Delle perdite di liquido possono danneggiare i componenti d d a a n n n n e e g g g g i i a a m m e e n n t t o o elettrici di questo strumento.
Page 68
All manuals and user guides at all-guides.com HEALTHCARE Installazione Contenuto Rete elettrica N.B. Caricare la batteria ... per almeno 12 ore prima della prima utilizzazione. durante tutta la notte se l'unità è utilizzata in modo batteria per più di 2 ore. almeno una volta ogni tre mesi se l'unità...
Page 69
All manuals and user guides at all-guides.com HEALTHCARE Collegamento del manicotto Inserire il tubo nel connettore. START Connettore del manicotto. Fissaggio della termosonda (SL510) Accensione Accendere Il LED si illumina Segnale sonoro L'unità è pronta START CYCLE STOP REVIEW START CYCLE CLEAR STOP...
Page 70
All manuals and user guides at all-guides.com HEALTHCARE Funzionamento Misurazione NIBP F F a a r r e e i i n n d d o o s s s s a a r r e e i i l l m m a a n n i i c c o o t t t t o o a a l l p p a a z z i i e e n n t t e e N N .
Page 71
All manuals and user guides at all-guides.com HEALTHCARE I passi 4 - 5 sono ripetuti 6 6 volte. Segnale sonoro. La misura media è indicata. START CYCLE Visualizzazione delle letture precedenti - per visualizzare la lettura precedente (il numero contrassegnato dalla dicitura "READING"...
Page 72
All manuals and user guides at all-guides.com HEALTHCARE Modo singolo N.B.: Per una misurazione singola, premere prima di applicare la sonda sulla parte del corpo prescelta. Accertarsi che la porzione metallica della sonda non sia a contatto con la pelle prima della misurazione. - Avvio della -Selezionare misurazione.
Page 73
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 74
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Allineare l'unità. Premere nella direzione del supporto, quindi HEALTHCARE verso il basso per fissarla. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Montaggio su supporto mobile (SL500-BAT e SL510) 1 1 x x...
Page 75
All manuals and user guides at all-guides.com HEALTHCARE Garanzia e riparazione I termini e le condizioni standard della divisione prodotti di Huntleigh Healthcare Diagnostic possono essere applicati a tutti i prodotti venduti. Copie sono disponibili su richiesta. Esse contengono tutti i dettagli dei termini di garanzia e non limitano i diritti del consumatore tali che definiti dalla legge.
Page 76
Alla ändringar och reparationer på å utrustningen ska utföras av kompetenta serviceingenjörer, å å terförsäljare eller sjukhustekniker som auktoriserats av Huntleigh Healthcare. 1. Varningar D D e e f f ö ö l l j j a a n n d d e e ä ä r r b b e e s s k k r r i i v v n n i i n n g g a a r r a a v v a a l l l l m m ä...
Page 77
All manuals and user guides at all-guides.com HEALTHCARE R R i i s s k k f f ö ö r r Sänk inte ner någon del av utrustningen I vatten. u u t t r r u u s s t t n n i i n n g g s s s s k k a a d d a a Vätskespill kan skada instrumentets elektriska komponenter.
Page 78
All manuals and user guides at all-guides.com HEALTHCARE Inställning Innehåll Nätström OBS: Ladda batteriet ... i minst 12 timmar före första användningen. över natten om apparaten används i batteriläget i mer än 2 timmar. minst en gång var tredje månad när den inte används, för att förlänga batteriets livstid.
Page 79
All manuals and user guides at all-guides.com HEALTHCARE Ansluta manschetten Tryck fast slangen START på anslutningen. Anslut till manschetten. Ansluta temperatursonden (SL510) Slå på strömmen Slå på Lysdioden tänds En ton ljuder Enheten är redo START CYCLE STOP REVIEW START CYCLE CLEAR REVIEW...
Page 80
All manuals and user guides at all-guides.com HEALTHCARE Drift NIBP-mätning S S ä ä t t t t p p å å m m a a n n s s c c h h e e t t t t e e n n p p å å p p a a t t i i e e n n t t e e n n . . O O b b s s : : a a v v s s l l u u t t a a r r m m ä...
Page 81
All manuals and user guides at all-guides.com HEALTHCARE 4 - 5 upprepas 6 6 gånger. En ton ljuder. Genomsnittlig mätning visas. START CYCLE Visa tidigare mätningar - visar tidigare mätning, (siffrorna på AVLÄSNING-skärmen motsvarar den mätning som visas). - raderar lagrade mätningar. Tryck på...
Page 82
All manuals and user guides at all-guides.com HEALTHCARE Enkelt läge OBS: För enstaka mätningar , tryck på innan sonden anbringas till mätningsstället. Se till att metalldelen i sonden inte kommer i kontakt med huden innan mätningen påbörjas. - starta -välj mätningen Programmet visar Enkel...
Page 83
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 84
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Rikta in enheten. Tryck mot konsolen och ner för HEALTHCARE att låsa. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Montering på...
Page 85
Märk paketet 'Service department - Smartsigns® LITE' Series'. Huntleigh Healthcare förbehåller sig rätten att returnera produkter som inte innehåller ett dekontamineringscertifikat. En servicemanual är tillgänglig för systemet och innehåller service information, reservdelslista och riktlinjer för felsökning.
Page 86
All manuals and user guides at all-guides.com HEALTHCARE Português Guia de Referência Rápido Página Conteúdo 1. Avisos ....... . . 86 2.
Page 87
All manuals and user guides at all-guides.com HEALTHCARE P P o o s s s s í í v v e e l l D D a a n n o o Não imergir partes do instrumento em nenhum tipo de n n o o E E q q u u i i p p a a m m e e n n t t o o líquido, pois pode danificar os componentes eléctricos...
Page 88
All manuals and user guides at all-guides.com HEALTHCARE Instalação Conteúdo Suprimento de Energia NOTA Carregue a bateria ... por 12 horas antes da utilização inicial. durante a noite se o instrumento for utilizado a bateria por mais de 2 horas. pelo menos uma vez a cada três meses quando não estiver em uso para prolongar a vida útil da carga da bateria.
Page 89
All manuals and user guides at all-guides.com HEALTHCARE Conexão de Manguito Conectar o tubo dentro do conector. START Conector ao manguito. Conexão da Sonda de Temperatura (SL510) Ligar Conectar LED iluminar-se-á Tom soará Unidade está pronta START CYCLE STOP REVIEW START CYCLE CLEAR...
Page 90
All manuals and user guides at all-guides.com HEALTHCARE Funcionamento Monitoramento de NIBP A A p p l l i i c c a a r r o o m m a a n n g g u u i i t t o o n n o o p p a a c c i i e e n n t t e e . . N N o o t t a a : : p p á...
Page 91
All manuals and user guides at all-guides.com HEALTHCARE As etapas 4 e 5 repetir-se-ão 6 6 vezes.Tom soará. A medição média é fornecida. Rever leituras anteriores START CYCLE - mostra a leitura anterior, (o número da LEITURA (READING) corresponderá a medição mostrada). - apaga as leituras armazenadas.
Page 92
All manuals and user guides at all-guides.com HEALTHCARE Modo Único NOTA: Para as medições únicas, carregue em antes de aplicar a sonda ao local de medição . Certifique-se de que a parte de metal da sonda não tenha contacto com a pele antes da medição. - Iniciar a -seleccionar medição...
Page 93
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 94
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Alinhe com a unidade. Prenda a unidade ao HEALTHCARE suporte da parede. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Instalação no Suporte Móvel (SL500-BAT &...
Page 95
Marcar no pacote 'Departamento de Técnica - Smartsigns® LITE' Series'. Huntleigh Healthcare reserva o direito de retornar produtos que não houverem um certificado de descontaminação. Um manual de serviço está disponível sobre o sistema e contém informações sobre assistência, listas de peças e linhas directrizes para...
Page 103
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 104
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . ÊñáôÞóôå ôç ìïíÜäá óå üñèéá èÝóç, ðéÝóôå ðñïò ôï õðïóôÞñéãìá êáé HEALTHCARE ôñáâÞîôå ðñïò ôá êÜôù ãéá íá áóöáëßóåôå. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Óå...
Page 106
All manuals and user guides at all-guides.com HEALTHCARE Nederlands Snelle Handleiding Pagina Nr. Inhoud 1. Waarschuwingen ......106 2.
Page 107
All manuals and user guides at all-guides.com HEALTHCARE De onderdelen van deze apparatuur mogen niet in water E E v v e e n n t t u u e e l l e e ondergedompeld worden. Gemorste vloeistoffen kunnen b b e e s s c c h h a a d d i i g g i i n n g g tot beschadiging van de elektrische onderdelen leiden.
Page 108
All manuals and user guides at all-guides.com HEALTHCARE Installatie Inhoud Netvoeding Laad de batterij op... ten minste 12 uur voordat u het systeem voor de eerste keer gebruikt. de nacht tevoren wanneer het systeem voor meer dan 2 uur in de batterijstand is gebruikt wanneer het systeem niet gebruikt wordt ten minste om de drie maanden om de levensduur van de batterij te verlengen.
Page 109
All manuals and user guides at all-guides.com HEALTHCARE De manchet aansluiten Duw de slang op START de connector Connector manchet Temperatuursonde bevestigen (SL510) Starten Zet het systeem aan Het LED gaat branden U hoort een pieptoon Het apparaat is klaar voor gebruik START CYCLE REVIEW...
Page 110
All manuals and user guides at all-guides.com HEALTHCARE Bediening NIBP Meting P P l l a a a a t t s s d d e e m m a a n n c c h h e e t t o o m m d d e e a a r r m m v v a a n n d d e e p p a a t t i i ë ë n n t t . . N N B B : : s s t t o o p p t t d d e e m m e e t t i i n n g g &...
Page 111
All manuals and user guides at all-guides.com HEALTHCARE 4 - 5 wordt 6 keer herhaald. De pieptoon gaat af. Gemiddeld meting wordt weergegeven START CYCLE Beoordelen van vorige aflezingen - geeft de vorige aflezing weer (het cijfer in het READING display komt overeen met de weergegeven meting).
Page 112
All manuals and user guides at all-guides.com HEALTHCARE Enkele functie NB: Druk tijdens het verrichten van enkele metingen op voordat u de sonde plaatst. Zorg ervoor dat het metalen gedeelte van de sonde niet in contact komt met de huid voordat de meting wordt verricht. - start de meting -selecteer De programma...
Page 113
Diagnostic Products Division 35 Portmanmoor Road Cardiff CF24 5HN HEALTHCARE Tel: +44 (0) 29 20485885 Fax: +44 (0) 29 20492520 Email: sales@huntleigh-diagnostics.co.uk Certified to: CAN/CSA C22.2 No. 601.1-M90 UL 2601-1:1988 202332 Complies with: SL500-B B AT T / SL510 4 4 x x...
Page 114
All manuals and user guides at all-guides.com HEALTHCARE 3 3 . . Zet het systeem op één lijn met de console en duw naar HEALTHCARE beneden om vast te zetten. Model Number: SL510 Caution: Subsitution of a component different from that supplied may result in measurement inaccuracies Montage op verrijdbaar statief (SL500-BAT &...
Page 115
De verpakking onder vermelding van 'Service Department - Smartsigns LITE' te retourneren. Huntleigh Diagnostics behoudt het recht om producten die niet voorzien zijn van een decontaminatiecertificaat aan de verzender terug te sturen. Er is een onderhoudshandleiding verkrijgbaar voor dit systeem. Deze handleiding bestaat uit onderhoudsinformatie, onderdelenlijsten en richtlijnen voor het verhelpen van storingen.
Page 116
Fax: +44 (0) 29 2049 2520 Email: sales@huntleigh-diagnostics.co.uk www.huntleigh-diagnostics.co.uk DISTRIBUTED IN DISTRIBUTED IN DISTRIBUTED IN THE AUSTRALIA BY: GERMANY BY: USA BY: Huntleigh Healthcare Pty Ltd Huntleigh Healthcare Inc HNE Healthcare Im Hülsenfeld 19 PO BOX 330 40 Christopher Way Hamilton Hill Eatontown 40721 Hilden...