• Under no circumstances hold the device or cu under water,
as this can cause liquid to enter and damage the device and
cu .
• If you store the device and cu , do not place heavy objects on
the device and cu . The cu line should not be bent sharply.
• Remove the batteries if the device is not going to be used for
a long period of time.
8. ACCESSORIES AND/OR REPLACE-
MENT PARTS
Accessories and/or replacement parts are available at
www.beurer.de, under "Service". Please state the corresponding
order number.
Designation
Universal cu (22-42 cm) 110.031
Mains part (EU)
Mains part (UK)
9. TROUBLESHOOTING
Error
Possible cause
message
Unable to record
a pulse.
The measured
blood pressure
is outside the
measurement
range.
Item number and/or order number
071.95
072.05
Solution
Please wait one minute and
repeat the measurement.
Ensure that you do not speak or
move during the measurement.
Error
Possible cause
message
The cu was
not attached
correctly.
An error occur-
red during the
measurement.
The infla-
tion pressure
is higher than
300 mmHg.
There is a
system error.
The batteries are
almost empty.
10. DISPOSAL
Repairing and disposing of the device
• Do not repair or modify the device yourself. Proper operation
can no longer be guaranteed in this case.
• Do not open the device except for the battery compartment.
Failure to comply will invalidate the warranty.
31
Solution
Please observe the information
in chapter "Attaching the cu ".
Please wait one minute and
repeat the measurement.
Ensure that you do not speak or
move during the measurement.
Please take another measure-
ment to check whether the cu
can be correctly inflated.
Make sure that neither your
arm nor other heavy objects are
pressing on the line, and that
the line is not bent.
If this error message appears,
please contact Customer
Services.
Insert new batteries into the
device.