The measured blood pressure value was obtained
when the user's circulatory system was su ciently
at rest and reliably represents the blood pressure
at rest.
Indicates that the value was obtained when the
user's circulatory system was not su ciently at
rest. The blood pressure values measured in this
case generally do not represent the blood pressure
at rest. The measurement should therefore be
repeated after a period of physical and mental rest
lasting at least 5 minutes.
No resting
During the measurement it was not possible to
indicator
determine whether the user's circulatory system
symbol is
was su ciently at rest. In this case too, the meas-
displayed
urement should be repeated after a period of rest
lasting at least 5 minutes.
The user's circulatory system not being su ciently at rest can
be the result of various factors, such as physical stress, mental
strain/distraction, speaking or experiencing cardiac arrhythmia
during the measurement.
In an overwhelming number of cases, HSD will give a very good
guide as to whether the user's circulatory system is rested when
a blood pressure measurement is taken.
However, certain patients su ering from cardiac arrhythmia or
chronic mental conditions may remain haemodynamically unsta-
ble even in the long-term, something which persists even after
repeated periods of rest. The accuracy of the results for the blood
pressure at rest is reduced in these users.
Like any medical measurement method, the precision of HSD is
limited and it can lead to incorrect results in some cases. Never-
theless, the blood pressure measurements taken when the user's
circulatory system is su ciently at rest represent particularly re-
liable results.
6.5 Displaying and deleting measured values
The results of every successful measurement are stored together
with the date and time. If there are more than 30 measurements,
the oldest measurements are lost.
• Press the memory button M. Select the desired user memory (
...
) by pressing the memory button M again.
• If you press the function button +, the average value
the stored measured values in the user memory will be dis-
played. If you press the function button + again, the average
value of the morning measurements for the last 7 days will
be displayed (morning: 5 am – 9 am, display
the function button + again, the average value of the evening
measurements for the last 7 days will be displayed (evening: 6
pm – 8 pm, display
button +, the most recent individual measured values are dis-
played in turn with the date and time.
• To switch o , press the START/STOP button
• If you forget to switch o the device, it will switch itself o
automatically after 30 seconds.
• If you want to delete the whole memory for a specific user,
press the memory button M. Press and hold the memory but-
ton M and the SET button simultaneously for 5 seconds.
7. CLEANING AND MAINTENANCE
• Clean the device and cu carefully using only a slightly damp
cloth.
• Do not use any cleaning solutions or solvents.
30
). If you press
). If you continue to press the function
.