Test the Hand Control
Perform a brief test to ensure proper connectivity:
1.
Turn on the lift as shown on page 7.
2.
Press any button on the hand controller. The hand controller screen lights up.
3.
Raise and lower the carry bar using the Up button
• If the carry bar moves up and down, the hand controller is properly connected.
• If the carry bar does not move correctly, double-check
That the lift is on
That the hand controller connector is full connected
That the hand controller does not display Slack Detected
If there are still problems with the lift, contact your local Handicare authorized dealer for
service.
Change the Hand Controller Hook
If you want to use a C450/C625 carry bar (360760) with your A-450/A-625 lift, Handicare recommends
you install an A-series hand controller hook (635669) on the hand controller for a more secure fit.
Turn On the Lift
Before using the A-450/A-625 lift system, check the lift, track, and sling
for any unusual wear or damage. Refer to the owner's manual for each
piece of additional equipment to determine what should be checked.
If anything looks unusual, contact your local Handicare dealer or your
local Handicare business development manager before use. Failure to
check the lift could result in damage to the lift or serious injury to the
operator or the individual being lifted.
1.
Press any button on the hand controller to turn the lift on. The indicator on the underside of the lift
lights green and the display screen on hand controller displays [Ready].
• If the lift batteries require charging, the indicator on the lift's underside lights orange, a beep
audible alarm sounds when the button is pressed, and the hand controller displays the message
[Low Battery].
• If the lift batteries are discharged to 25-30% of capacity, the hand controller displays the message
[Low Battery] and the lift continues to function until the battery status reaches 10-15%.
• If the lift batteries are discharged to 10-15% of capacity, the hand controller displays the message
[Very low battery], the indicator on the lift's underside lights red, and a beep audible alarm sounds
when the button is pressed. The lift will not raise, but will lower and the emergency lowering will
still function.
Turn Off the Lift
The lift automatically shuts off after approximately 2 minutes of non-use.
In an emergency, the lift can also be shut off by pulling the Emergency Down strap. See
"Emergency Stop" (page 13) for details on Emergency Stop and restarting the lift after an
Emergency Stop.
Hand Controller Operation
The A-450/A-625 lift hand control
Controls lift motion
Presents active lift status display
Provides usage data and diagnostics.
and Down button
7
.

Publicité

loading

Ce manuel est également adapté pour:

A-625