Verify all functions on the hand controller work correctly (Up, Down, Traversing, Return to Charge,
etc.).
The lift makes no unusual sounds when the carry bar is moved up or down or the lift is moved left
or right.
Press and hold the Up button until the carry bar is at the top. The motor should stop
automatically.
Using the hand controller, let the strap out all the way to the ground. The motor should stop
before the strap begins to wind backwards.
While using either the up or down buttons, angle the strap more than 15 degrees. The motor
should stop automatically.
Engage the unit with charger and check that the unit is charging properly. The indicator on the lift
should light orange.
Check that the lift's LED light lights green when on, dark when off, and orange when charging
Check for cracks or wear in the case.
If the audible/visual preventive maintenance alert sounds, complete the Preventive
Maintenance Form before resetting the lift counter. Instructions to reset the lift counter
are detailed in the A-Series Technical Manaul. Counter can only be reset by an authorized
Handicare dealer.
Annually
(To be completed by a lift technician)
This section to be only completed by a qualified service technician as authorized by
Handicare.
Annually, the A-450/A-625 system must be inspected. Be sure to documents this service history as
directed by the A-series Tech Manual.
Complete the inspection as noted in the "Monthly" section above.
Recurrent Performance and Safety Testing
The following table highlights A-450/A-625's Essential Performance Parameters, with their subsequent
testing method and frequency. If your A-450/A-625 does not meet the following criteria, contact your
local Handicare authorized dealer for service.
Performance Criteria
The A-450/A-625 should be
capable of lifting 450 lb /
204.5 kg (for A-450) or 625 lb
/ 284 kg (for A-625)
The A-450/A-625 will
operate as intended
throughout its service life,
provided all recommended
maintenance and usage
recommendations are
followed.
Recommended Check
Load test at full load (625 lb /
284 kg for A-625 and 450 lb /
204.5 kg for A-450).
Load test at full load (625 lb /
284 kg for A-625 and 450 lb /
204.5 kg for A-450)
Pre-lifting inspection by user
caregiver.
Preventive maintenance by
trained tech.
26
Frequency
Annual
Annual.
Every use
Semi-annually or annually, per usage
frequency.
See "
General Inspection and Preventive
Maintenance" on page 24.