Fault
One side of the lift strap is
starting to fray after
continued use.
The lift does not pass
through a track component
such as a turntable or
transistion gate.
No Power
Lift starts beeping while
pulling Emergency Down
Strap/Switch with strap not
coming down.
The user weight readings
fluctuate too much while
weighing a patient on a load
cell equipped lift.
The user weight readings are
not accurate while weighing
a patient on a load cell
equipped lift.
The lift is recommending
preventive maintenance.
Discontinue using immediately and replace strap. Check to be sure that
the lift is always directly above the individual being lifted, especially with
power traversing lifts. Refer to "Transfer an Individual" on page 9 for
correct lift positioning. If fraying still continues, contact your local
authorized dealer immediately to check the lift and ensure proper
continued operation.
Refer to the Owner's Manual for the specific piece of equipment in
question. If the recommended solution does not correct the problem,
contact your local authorized dealer immediately to check the lift and
ensure proper continued operation.
Ensure the Emergency Lowering tab has not come out. If it has, simply
press the tab back into the lift and check battery for charge.
The A-series lift is equipped with dual sensors for strap tension and
position, the Emergency Down Strap/Switch system only works with load
on the strap and Emergency Down Strap/Switch motion stop immediately
after the lift strap is loose.
Ensure the user is stable and there is no swinging movement while
weighing a patient.
If problem persists, contact your local authorized dealer immediately so
that the lift can be checked to ensure proper continued operation.
Ensure the user is stable and there is no swinging movement while
weighing a patient. Also ensure the scale is properly tared/zeroed as
described in "Weigh an Individual" on page15. If the problem persists,
contact your local authorized dealer immediately so that the lift can be
checked to ensure proper continued operation.
Complete the "Semi-Annually" maintenance procedure on page 24 prior
to resetting the lift counter. Instructions to reset the lift counter are
detailed in the A-Series Technical Manaul. Counter can only be reset by an
authorized Handicare dealer.