Télécharger Imprimer la page

Handicare A-450 Mode D'emploi page 27

Publicité

Performance Criteria
The A-450/A-625 should be
capable of working in
conjunction and
compatibility with
Handicare/Prism Track,
trollies and accessories
(Carry bars and Slings).
The A-450/A-625 should
have "Free Fall" protection
system.
The A-450/A-625 should
have upper and lower limits.
The A-450/A-625 strap
should have a minimum
safety factor of 6 in terms of
break load.
The A-450/A-625 should
have visual "Error" status
display.
The A-450/A-625 hand
control locking mechanism
used as connections should
be "Self-Locking".
Recommended Check
Load test at full load (625 lb /
284 kg for A-625 and 450 lb /
204.5 kg for A-450).
Pre-lifting inspection by user
caregiver.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
Load test at full load (625 lb /
284 kg for A-625 and 450 lb /
204.5 kg for A-450).
Pre-lifting inspection by user
caregiver.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
Preventive maintenance by
trained tech.
27
Frequency
Annual
Every use
Semi-annually or annually, per usage
frequency
See "
General Inspection and Preventive
Maintenance" on page 24.
Semi-annually or annually, per usage
frequency. See "
General Inspection and Preventive
Maintenance" on page 24.
Semi-annually or annually, per usage
frequency. See "
General Inspection and Preventive
Maintenance" on page 24.
Annual
Every use
Semi-annually or annually, per usage
frequency
See "
General Inspection and Preventive
Maintenance" on page 24.
Semi-annually or annually, per usage
frequency.
See "
General Inspection and Preventive
Maintenance
" on page 24.
Semi-annually or annually, per usage
frequency.
See "
General Inspection and Preventive
Maintenance
" on page 24.

Publicité

loading

Ce manuel est également adapté pour:

A-625