4. Emergency Stop
The red Emergency Stop Button is located on the front of the control box and is acti-
vated by pressing in. This will cut all power to the lift and only be reset by twisting
the button anticlockwise and releasing.
5. Electrical Emergency Down/Up
Emergency lower (down) and raise (up) buttons are provided at the front of the control
box. This is operated by inserting a ballpoint pen into the button highlighted
.
This will bring the boom up or down should the hand control fail at any time.
6. Mechanical Emergency Down
In the case of a complete electrical failure the Stature lift is fitted with mechanical
lowering device. This device is located on the left hand side of the lift's chassis.
The lowering mechanism is operated by inserting a 6mm Hex socket into the aperture
and onto the mechanism turning clockwise to raise the lift and anticlockwise to lower
the lift. This can be done manually with a socket wrench.
Recommendation!
In order to raise or lower the boom more quickly it is recommended that a 6mm Hex
socket be used in conjunction with a 1/4 inch drive adaptor and an electric reversible
drill. (Note– do not use hammer setting.) This will increase the speed in which the
patient can be safely raised and lowered. A 6mm Hex socket is contained in the top cap
located on the top of the mast.
294000.10017 Rev. B