Test ID
Testsubject
1
Max. Deviation to
Mean/well on
sample tracking
system / tube
holder
2
Min.
Detectionposition/
well sample
tracking system /
tube holder
3
Max. Deviation to
Mean/well on S-
Block
4
Min.
Detectionposition/
well on S-Block
(except empty
wells)
5
Max.
Detectionposition/
well on S-Block
(empty wells)
MultiFlow Dispenser, Dispense Test results:
Position: Therm Slot / Adapter (RB) / S-Block
(66) Flow rate of the MultiFlow Dispense Pump in the Robot Configuration: 412.00 ul/s.
Testsubject
Max. Accuracy / A3|H8
Max. CV / A3|H8
BioRobot: BioRobot_MDx_DSP_SNxxxx
Robot Configuration: BIOROBOT_MDX_DSP
QIAsoft: MDx DSP 1.0.0.5
Message Log File: 040908AB.MSL
Test Protocol: SOW-00000136-000-A_BR MDx DSP Maintenance Protocol.PRO
Date/Time: 08.Sep.2004, 15:49
OPERATOR: generaloperator , NAME: ______________________ ,
SIGN: ______________________
BioRobot MDx DSP Manuel d'utilisation 10/2004
Limit
Limit
min.[steps]
max.[steps]
50
50
Limit max.[%] Measured[%]
12
5
Measured
[steps]
30
11.00
587.09
30
12.00
628.00
50
0.00
Result
4.32292
3.43979
Index
Result
PASSED
PASSED
PASSED
PASSED
PASSED
PASSED
PASSED
Index-5