Configuration
1) Notification to CMS: Switch for
outbounding data to CMS, default:
unchecked.
2) Notification Type: Data notification to CMS
through auto retrieving or manual inputting
IP address, default: Auto.
Under situation that Notification to CMS is
checked, if the notification type is Auto,
Gateway automatically retrieves the CMS
connected to Gateway and transfers the data to
the CMS according to its requirement; if the
notification type is Manual, the user should
manually input the IP address of CMS to be
notified.
Realtime Data Config: Gateway outbounds
device data in real-time to EMR periodically.
The central station outbounds periodical data
as per its own period.
1) Periodic Outbound Data: switch for
Gateway outbounding periodic data to
EMR, default: checked.
2) Periodic Data Contains NIBP: switch for
periodic data containing NIBP, default:
checked.
Time Synchronize
The user shall synchronize the time of Gateway
and external system before Gateway is connected
to the external system. The time synchronization is
performed in the Time Synchronization interface.
48
3) Sending interval: interval time for periodic
data outbounding, deault: 60s.
4) Sending NIBP after measurement: switch
for outbounding NIBP unsolicitedly, default:
checked.
5) Sending QuickTemp: switch for
outbounding QuickTemp unsolicitedly,
default: checked.
Alarm Switch Config: the configuration items
for Gateway outbounding alarms to EMR.
1) Parameter Alarm: switch for outbounding
parameter alarm message, default:
unchecked.
2) Arrhythmia Alarm: switch for outbounding
arrhythmia alarm message, default:
unchecked.
3) Life Alarm: switch for outbounding Life-
threatening alarm message, default:
unchecked.
Device Query Config: the user can configure
Query ID, default: Device GUID. Gateway
queries the devices through Device SN, Device
No. and Device GUID.
Click Refresh to refresh the current configuration
information. Click Save to save the current
modification.
Vista Gateway Instructions for use