17. International regulations
17.1 Electromagnetic compatibility (EMC)
The Swing Maxi
™
Hands-free breast pump complies with all requirements
of the relevant standards and regulations with regard to electromagnetic
emission and immunity to interference.
17.2 Simplified EU Declaration of Conformity
Hereby, Medela declares that the radio equipment type Swing Maxi
Hands-free is in compliance with Directive 2014/53/EU. The full text of the
EU declaration of conformity is available at the following internet address:
https://www.medela.com/eu-conformity-swing-maxi-hands-free
18. Technical specifications
Summary of important technical specifications
Vacuum performance (at 500 m above sea level [954.62 hPa ambient pressure])
-45 to -245 mmHg (-60 to -327 hPa)
45 to 111 cpm
Power adapter (Model: MS-V2000U050-010B0-* / MS-H2000U050-010B0-*)
*country code
Input
100 to 240 V~ 50/60 Hz max. 0.5 A
Output
5 V
Average efficiency
78.7 %
No-load power
less than 0.075W
consumption
Motor unit
Battery capacity and type: 3.6 V, 2500 mAh (nominal) Li-ion.
Size: 126 x 125 x 58 mm. Weight: 325 g. Ingress protection level: IP22.
Transmission standard: Bluetooth
Frequency range: 2402-2480 MHz
Max. power output: +4 dBm (class 2)
2 A
10 W
version 4.2 BLE
®
WARNING
The Swing Maxi
Hands-free breast pump should not be used adjacent to or
™
stacked with other equipment. If adjacent or stacked use is necessary, the Swing
Maxi
Hands-free breast pump should be observed to verify normal operation
™
in the configuration in which it will be used.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations, walkie-talkies can
affect the electric breast pump and should be kept at a distance of at least
30 cm away from the device.
™
Use of accessories or cables other than those provided by the manufacturer of
this device could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
+35ºC
Materials touching skin or coming in contact with milk
• Breast shield : Polypropylene
• Outer shell : Polypropylene, Thermoplastic Elastomer
• Membrane : Silicone
• Tubing : Polyvinyl Chloride, Acrylonitrile Butadiene Styrene
Not intentionally made of BPA.
Operating temperature range
(5 °C to 35 °C)
(41 °F to 95 °F)
Operating humidity range
(15% to 93% relative humidity)
Operating atmospheric pressure range:
70kPa...106kPa
Transport/storage
temperature range
(-20°C to 50°C)
(-4 °F to 122 °F)
Transport/storage
humidity range
(15% to 93% relative humidity)
21