Instructions for use for medical bariatric work chairs
ID
Area
11 Accessories
11.1 Restoration
11.2 Mechanics
11.3 Function
12 Final inspection
12.1 Final inspection
12.2 Battery sleep mode
Reconditioning instructions
Remove the accessories that should not be attached to the
device in the basic version, see 5.2. Basic design.
Check all welds and that there is no damage to mechanical
parts that may impair or weaken the structure. Ensure
good lighting and preferably use a magnifying glass, as
small cracks can be difficult to detect on black lacquered
structures.
Check and replace any worn parts. Check that the locking
and function of moving parts works properly.
Test all functions of the device. Check that all settings,
electrical functions and brakes are working properly. Check
that the castors are working properly and that all four
wheels are in contact with the surface.
If electrical models are to be stocked, the battery can be
put into sleep mode with the intended 9V adapter, see
8.3. Reconditioning and periodic maintenance, Long-term
storage of batteries.
45
Servicing & reconditioning
21-08861-EN-04