Instructions for use for medical bariatric work chairs
Warning! Electric models have lifting
mechanisms that can lift the user
so high that this might pose a risk.
If the user cannot get out of the
device from its highest position,
then the seat height should never
be higher than the position at which
the patient's feet touch the ground.
The only exception to this is cases
where the user can summon help via
a home-help alarm or telephone.
Warning! Only batteries and chargers
from Mercado Medic that have
been tested and approved for use
with the product may be used with
models with electrical functions.
If using other chargers or batteries,
the CE marking is not applicable and
Mercado Medic AB's device liability
will cease to be valid for any cases
regarding the battery, charging or
other electronics.
Warning! The device must not be
equipped with any accessories
or components other than those
approved by Mercado Medic.
To maintain the CE marking, no
individual modifications or changes
may be carried out without Mercado
Medic's approval. See section 5.12 if
a non-approved component it to be
used. individual customisation.
Warning! Do not plug any electrical
accessory, other than those
customised by Mercado Medic,
into the control system's contacts.
Non-adapted components are at
risk of breaking and in turn injuring
the user. See section 5.12 if a non-
approved component it to be used.
individual customisation.
Warning! Installation, connection
or dismantling is not risk-free.
If components are handled
incorrectly, for example, crushing
damage may occur. This type of
work may therefore only be carried
out by a Mercado Medic authorised
technician.
Warning! All parts of the device must
be installed and fixed so that there is
no risk of small parts coming loose.
All cables must be secured with
cable ties to the device's structure to
minimise the risk of strangulation.
6.3. Expected service life
The expected service life of the device is ten
years when used in accordance with these
instructions for use. The expected service
life of the device is calculated from the date
of manufacture of the device. The date of
manufacture of the device can be found
on one of the silver labels on the device's
base and is in the format YYYY-MM (year
and month), see Figure 6.1. The label also
includes serial numbers and an identifier for
the device model (UDI-DI).
REAL 9200 TWIN
Basic UDI-DI: 732184D0029R
Mercado
Medic AB
Tryffelslingan 14
SE - 181 57 Lidingö
www.mercado.se
Identifi cation bar codes:
00123456700012345600123456701PV0000
Example label 123
Figure 6.1. Serial number, UDI-DI and date of manufacture on the
label.
29
Important safety information
180 kg
###### 001
90######AB
YYYY-MM
2000m
21-08861-EN-04