Control drawing A5E52688321A/001
Control drawing
Sensor
Notes
i. Multi-core cables connecting separated intrinsically circuit D+/D-, S1+/S1-, S2+/S2- and TP1/TP2/TP3 shall
have suitable insulation as required by NEC 504.30(B).
ii. Screens of multi-core cable for individual intrinsically circuit shall be connected with COM terminal on
remote type converter.
iii. Hazardous location depends on order code (MLFB):
(j = H): Transmitter: Class I, II, III, Div. 1, Gr. A, B, C, D, E, F, G
Sensor:
(j = J):
Transmitter: Class I, II, III, Div. 1, Gr. C, D, E, F, G
Sensor:
iv. WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
v. WARNING: POTENTIAL ELECTROSTATIC CHARGING HAZARD
Transmitter
or US: Class I, Zone 1, Gr. IIC; CAN: Zone 1 per CEC 18-100 & 18-090
Class I, II, III, Div. 1, Gr. A, B, C, D, E, F, G
or
US: Class I, Zone 0, Gr. IIC
CAN: Ex ia IIC Ga
or US: Class I, Zone 1, Gr. IIB; CAN: Zone 1 per CEC 18-100 & 18-090
Class I, II, III, Div. 1, Gr. C, D, E, F, G
or
US: Class I, Zone 0, Gr. IIB
CAN: Ex ia IIB Ga
16