Monthly, the AP300/450 lift and associated track, accessories, and slings, must be inspected.
Should any of these items fail the inspection do not use the lift. Contact your local
authorized dealer for service.
Complete the visual inspection as noted in the "Each Use" section above.
With no one in the sling nor attached to the lift check that the lift moves freely along the
entire length of the track.
Semi-Annually
(To be completed by User)
In high frequency transfer areas (more than 1500 lifts per year) or situations where heavier than
normal clients are regularly lifted, maintenance should also be completed every 6 months. It is
recommended that this service history be documented.
Should any of these items fail the inspection, do not use the lift. Contact
your local authorized dealer for service.
Check for the following:
Check emergency stopping and emergency lowering functions using the red Emergency
Stop Switch and Emergency Down button.
Inspect lift hooks and exterior for damage Ensure there are NO cuts, dents, or sharp
edges that may damage the strap of the unit or sling.
Lower the lift exposing the entire length of strap; verify there are no signs of fraying or
breaking along its entire length.
The stitching on the lift strap where it connects to the carabiner shows NO signs of fraying
or breaking.
Verify all functions on the Hand Control work correctly (Up, Down, etc.).
Verify all functions on Control Membrane (Up, Down, Emergency Down)
The lift makes no unusual sounds when moving up, down, left or right
Press and hold the Up button until the lift is at the top. The motor should stop
automatically.
Using the Hand Control, let the strap out all the way to the ground. The motor should stop
before the strap begins to wind backwards.
While using either the up or down buttons, angle the strap more than 20 degrees. The
motor should stop automatically.
Engage the unit with charger and check that the unit is charging properly. The indicator on
the lift should light orange.
Check that the lift's LED light is green when on, dark when off, and orange when charging
Check for cracks or wear in the case.
If the audible/visual preventive maintenance alert sounds, complete the
Preventive Maintenance Form before resetting the lift counter. Instructions to
reset the lift counter are detailed in the AP300/450 Service/Spare Parts
Manual.
Annually
(To be completed by a lift technician)
This section to be only completed by a qualified service technician as
authorized by Handicare.
Annually, the AP300/450 system must be inspected. Be sure to documents this service history.
Complete the inspection as noted in the "Monthly" section above.
Complete the preventive maintenance procedure as outlined in the AP300/450 Technical