Hand Control Information and
In addition to controlling the lift, the AP300/450 Hand Control is an interface that provides usage
and diagnostics data from the lift.
To view data and setting through the Hand Control:
1.
Press the
2.
Use the
3.
The Scale menu is used on a feature that is not available for AP300/450.
4.
The Usage Stats menu is for reviewing usage information for the lift.
5.
The Diagnostics menu is for service and technical support of the lift.
6.
On the menu you want to review, press the
7.
If you select the wrong menu, at the bottom of the display:
•
Select EXIT/GO BACK and press the
or
•
Press the Up
The lift unit also has an Emergency Shut-off feature so that in an emergency the operator can
completely cut the power to the lift.
To perform an Emergency Stop:
1.
Press the Emergency Stop Switch located on the back side of the AP300/450 to OFF.
2.
The lift immediately stops, and all power to the lift shuts off
•
To lower the patient after an emergency stop, press down on the Emergency
Down button located on the lift's membrane.
3.
All lift functions are disabled
4.
The On indicator light and display turn off
Should the emergency stop function be used in an emergency, it must be inspected by an
authorized dealer. Once the lift has been inspected and/or repaired, the lift can be turned back
on. To turn the lift on after an Emergency Stop:
1.
Inspect the lift.
2.
Press the Emergency Stop Switch located on the back side of the AP300/450 to ON.
The Emergency Stop Switch must be pushed back into "ON" in order to restore power
to the lift.
The Emergency Lowering System can be used to lower the lift should the Down button on the
hand control and lift membrane fail. When pressed, an audible alarm will sound until the
Emergency Down button is released. When pressed, the Emergency Down Button bipasses all
other control circuitry of the lift to immediately lower an individual safely. To perform Emergency
Lowering:
Locate the Emergency Down button on the lift's membrane (Figure ).
1.
2.
Press and hold the Emergency Down button until the lift and individual are lowered to a
safe location.
Diagnostics
key of the Hand Control. The Menu screen displays.
Figure 12: The Hand Control Menu Screen
keys to scroll through the menu options.
or Down
Emergency Stop
Emergency Lowering
key. The selected menu displays.
key to return to the previous screen
lift buttons to return to the main screen.
14