17 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
•
Prepare the required adjustment weight. The adjustment weight to be
applied depends on the capacity of a weighing scale, see chap. 1. Carry
out adjustment as closely as possible to admissible maximum load of
weighing scale. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com.
•
Observe stable environmental conditions.
For warm-up time required for stabilisation see chap.1.
Procedure:
In weighing mode Actuate adjustment switch. The service
Press
Acknowledge by
44
menu is called up. The first menu item "X10" appears.
, the next menu item "AdJuSt" appears
„cAL" is displayed.
TMWN_A-BAPS-def-2211