10 Menu
Access to service menu „x10" is locked in verified balances.
To disable the access lock, destroy the seal mark and actuate the
adjustment switch.
Attention:
After destruction of the seal the weighing system must be re-verified by
an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.
10.1 Navigation in the menu
Call up menu
Select function
Change settings
Exit menu/
Return to weighing
mode
30
 In weighing mode press
„SetuP" followed by „coM" will appear.
 With help of
can be selected one after the other.
 Confirm selected function by
will be displayed.
 Select desired setting with
, the balance returns to the menu.
 Press
repeatedly until the zero display appears.
The balance is now in weighing mode.
simultaneously,
or
, the individual functions
. The current setting
and confirm by