During operation, the following environmental conditions must be
respected:
In the presence of magnetic fields the pump must be protected using
a ferromagnetic shield. See the appendix "Technical Information" for
detailed information.
The TV 401/301 MS must be connected to a primary pump (see
"Technical Information").
The TV 401/301 MS can be installed in any position. Fix the TV
401/301 MS in a stable position connecting the turbopump fixing
flange to a fixed counter-flange capable of withstanding a torque of
2000 Nm around its axis.
The following table shows the screws to be used to fix the
turbopump, the necessary number and the relevant fixing torque.
Tab. 1
VACUUM CHAMBER FIXING
M6 screws, Resistance Class >= 10,9
WARNING!
The vacuum chamber that contains the Turbopump must have an adequate
thickness for safety requirements. For example, a thickness of 5 mm is
adequate for an hardened Anticorodal chamber (σ
TV 401/301 MS User Manual / 87-901-013-01
temperature: from +5 °C to +35 °C (see the diagram pressure-
temperature in the appendix "Technical Information")
relative humidity: 0 – 95 % (non-condensing)
y
(A)
Instructions for Use
Installation
N.
FIXING TORQUE
6
10 Nm
=275 MPa).
199/264
16