K. Troubleshooting
LED
Color / Status
No lights
PP and S Green / Solid
PP and S Orange / Flashing
alternating
PP and S Orange / Flashing
simultaneously
PP and S Orange / Flashing
simultaneously
PP and S Orange / Flashing
simultaneously
Beeps
Condition
None
No probe loaded or driver
in sleep mode if not moved
for 120 seconds.
1 beep
Probe loaded in driver.
(only when
loading probe)
None
Sample container
removed.
2 beeps
Tissue sample stuck in
sample chamber or probe
chamber is open and
probe is stuck in driver.
Up to 9 beeps
Keypad button held
down after completing
prime, pierce, or sample
acquisition sequence.
5 beeps
Probe inserted before
driver has reset.
Action Required
Move driver or place driver on charging
stand. PP and S lights will turn solid green.
after loading a probe into the driver,
ensure the probe cover and locking
tab are securely in place and ensure
the probe is new and has never been
used before.
remove the probe and check to see
pressing and holding the S button. If
the Battery indicator light indicates
with a new probe.
after properly inserting a new probe,
contact Bard Service and Repair.
See section J, Driver Charging, for Battery
light and charging stand instructions.
Driver ready to perform biopsy. Option to
either "Prime" or "Sample."
Ensure sample container is installed in
probe.
If procedure is not yet complete, detach
coaxial from probe to maintain location in
breast. Remove probe from the breast.
PP and S buttons simultaneously for
at least 3 seconds until the chamber
opens. Remove any tissue from
sample chamber and then press and
hold the PP and S buttons simultane-
ously for at least 3 seconds. If both
the PP and S lights are solid green,
the biopsy can be continued. Insert
the probe and reattach the coaxial to
continue the biopsy.
any tissue from the chamber and then
press and hold the PP and S buttons
simultaneously for at least 3 seconds.
If both the PP and S lights are solid
green, the biopsy can be continued.
holding the PP and S buttons simultaneously
for at least 3 seconds, remove the biopsy
probe, reset the driver, and replace with a
new probe if additional samples are required.
Release the keypad button. Either PP or PP
and S button(s) will turn solid green, and the
biopsy procedure can be continued.
Remove probe from driver, allow driver to
reset and reload probe after driver has reset.
7