Troubleshooting the Purely Yours Ultra Breast Pump
To check pump suction from motor:
• Remove tubing adapter from tubing adapter port of pump.
• Put a finger over tubing adapter port.
• If suction is felt, pump motor is working properly.
• If suction is not felt, unplug AC adapter, wait a few seconds, plug back into electrical outlet.
The Purely Yours Ultra Breast Pump has a reset feature. If the Purely Yours Ultra Breast Pump
motor stops running:
• Unplug AC adapter cord from electrical outlet.
• Wait a few seconds and then plug AC adapter back into outlet.
Troubleshooting the HygieniKit Milk Collection System
If milk is dripping too slowly into bottle:
• Set SPEED dial to a faster setting.
• Replace silicone diaphragm(s) and white valve(s).
If pumping causes discomfort:
• Set pump to lower SUCTION setting.
• Check breast flange fit. A larger/smaller breast flange or insert may be needed.
If suction is low:
• Check that white valve(s) is/are pushed onto breast flange(s).
• Check white valve(s) for tears or stretches. Replace if needed.
• If single pumping, remove one tube and close one side of the tubing adapter.
• Check tubing adapter for cracks. If seen, replace.
• Set pump to higher SUCTION setting.
The white valves are vital for proper suction. To work well, they must be intact and look closed
when pump is not in use. To prolong white valve life, follow cleaning instructions.
9
E N G L I S H