Cleaning the Ameda Elite Breast Pump
Troubleshooting
Pump Does Not Turn ON
• Ensure proper power cord connection to pump.
• Ensure power cord connection to wall outlet.
• Ensure outlet has power.
Low or No Suction
• Check breast flange for adequate seal against breast.
• Check that the Ameda HygieniKit Milk Collection System
is properly assembled.
• Check that adapter cap is firmly snapped onto top of
breast flange.
• Ensure silicone diaphragm is in place inside top of breast
flange.
• Ensure tubing adapter is properly fitted into tubing
adapter port on pump.
• Inspect valve for stretching or tears. If seen, replace
valve.
• Ensure valve is pushed firmly into place in lower portion
of breast flange.
• Ensure one side of tubing adapter is closed when single
pumping.
Maintenance
It is recommended the Ameda Elite Breast Pump be
inspected at least once every year to include the following:
• Visually check pump housing for cracks, breakage, or
damage and ensure pump feet are present.
• Visually check for power cord damage.
• Check for proper operation of controls and buttons.
• Check there are no missing screws.
• Disconnect the pump
power cord from wall
outlet.
• Remove piston cylinder by
placing index finger into
the tube, lifting upwards
and turning right to release
from attachment bars.
For At-Home/End-Users
After each pumping, wipe
pump with cloth dampened
with warm, soapy water.
For Hospitals/Rental Stations
Clean between users.
Spray with recommended
disinfectant, such as
CaviCide
. Wipe with soft
TM
cloth. Do not use products
containing phenol.
Do not use bleach on pump.
• Place piston cylinder back
onto attachment bars and
press gently into place with
thumb.
• Check tubing adapter port on piston/cylinder is free from
obstruction.
Quick Functional Check
The following checks/tests should be performed prior to use
by a new user:
• Visual checks for power cord damage.
• Check for abnormal noise.
• Check for proper operation of controls.
Replacing Detachable Power Cord (if applicable)
• Power cord is detachable from power inlet.
• If power cord condition is in doubt, only replace with
Ameda supplied 2 conductor cord.
Technical Information/ Icon Glossary
Main voltage
Elite with Detachable Power Cord 100VAC to 240VAC
Elite with Built-In Power Cord 100/120V ~ or 220/240V ~ /AC
Frequency
50 / 60 Hz
Power consumption
30 VA
Main fuse
T 1.6A
Protection class
Class II
Safety level
Drip protection
IP 21
Consult Instructions for Use
Weight
6.4 lbs/2.9 kg
Dimensions
10.0"L x 9.0"W x 7.5"H
22.2cmL x 26.7cmW x 18.4cmH
Vacuum max.
[mbar]
Single HygieniKit
350 ( -10% )
Dual HygieniKit
310 ( -15% )
Variable cycle
30 - 60 / Min
Conforming to
E325189 MEDICAL EQUIPMENT
WITH RESPECT TO ELECTRIC SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY.
IN ACCORDANCE WITH
IEC60601-1: 1988 + A1:1991 + A2:1995,
IEC60601-1-2, CAN/CSA C22.2 No. 601.1-M90
!
Caution Consult Documents
Waste Electrical and
Electronic Equipment
minimum
size
Fragile
Protect from sunlight
Keep Dry
130ºF
Temperature Limitation For Transport/Storage
54ºC
-10ºF
-23ºC
90
Humidity Limitation For Transport/Storage
%
20
93ºF
Temperature Limitation For Operation
34ºC
53ºF
12ºC
70
Humidity Limitation For Operation
%
20
Double isolated
= BF
[mm Hg]
262 ( -10% )
232 ( -15% )
Manufacturer
SN
Serial Number
LOT
Lot Number
REF
Catalogue Number
E N G L I S H
7