8.2
Restore factory setting
NOTICE!
For the ErgoOne® E Multi-Channel pipettes, a volume offset is applied during adjustment (i.e.
the volume changes by the same amount over the entire volume range of the pipette). It is
recommended to perform the adjustment at 50% of the nominal volume.
NOTICE!
The instrument is permanently adjusted for aqueous solutions but can also be adjusted for
solutions of different density, viscosity, and temperature. The ErgoOne® E Multi-Channel pipettes
can be adjusted in any mode (except GEL mode).
9.
Disinfection/autoclaving
9.1
UV sterilization
The device is resistant to normal exposure to a UV disinfection lamp. The effects of the UV
exposure may cause some color change.
24 · ErgoOne
E Multi-Channel Pipettes · User Manual
®
The CAL symbol permanently shown in the display
indicates that an adjustment has been made.
a.
1x
>3s
b.
1x
c.
Enter CAL mode: Pressing and holding (> 3 s) the
menu key invokes the CAL mode.
The display shows 'on'.
'CAL' flashes.
Switch off CAL mode: Pressing one of the arrow
keys deactivates the CAL mode.
The display changes from 'on' to 'off'.
'CAL' continues to flash.
Restore factory condition: Press 'Enter' key.
The permanently displayed CAL icon has
disappeared.
The instrument has been returned to the original
factory condition.