5.3
revPIP mode
Program especially for pipetting liquids with high viscosity, high vapor pressure, or foaming media.
Volume and speed setting, see Set volume and Set the aspiration and dispensing speed.
16 · ErgoOne
E Multi-Channel Pipettes · User Manual
®
1x
c. Confirm PIPmix mode: Press 'Enter' key.
The display now shows 'blo' for blow-out.
d.
Preparing for pipetting: Press the pipetting key
once to move the piston to its start position.
The arrow in the display points upwards
(aspiration).
1x
e.
Aspirate liquid: Press the pipetting key once
to aspirate the liquid.
1x
f.
Dispense liquid in PIPmix mode: Pressing and
holding the pipetting key causes the liquid to
be alternately dispensed and aspirated.
The display alternately shows the arrow
symbol for aspiration or dispensing as well as the
Hold pressed
number of cycles.
g.
End pipetting: Pressing the pipetting key once
dispenses the liquid and triggers the blow-out.
After dispensing the residual liquid (blow-out), the
display jumps back to the set mode (start position).
1x
NOTICE!
The display shows a maximum of 19 cycles.
3x
a.
Call up menu selection: Pressing the menu key
three times takes you to the program selection.
'Mode' flashes.