StarLab ErgoOne E Mode D'emploi page 18

Masquer les pouces Voir aussi pour ErgoOne E:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

5.4
Electrophoresis (GEL) mode
Program for loading electrophoresis gels. A previously defined sample volume is aspirated at a
high, variable speed and slowly released again.
Volume and speed setting, see Set volume and Set the aspiration and dispensing speed.
18 · ErgoOne
E Multi-Channel Pipettes · User Manual
®
i.
End pipetting: By pressing the pipetting key once,
the blow-out is triggered, and the residual liquid is
dispensed.
After dispensing the residual liquid (blow-out), the
1x
display jumps back to the set mode (start position).
3x
a.
Call up menu selection: Pressing the menu key
three times takes you to the program selection.
'Mode' flashes.
b.
Set GEL mode: Use one of the arrow keys to scroll
through the modes until 'GEL' appears.
1x
c. Confirm GEL mode: Press 'Enter' key. The display
now shows 'blo' for blow-out.
d.
Preparing for pipetting: Press the pipetting key
once to move the piston to its start position.
The arrow in the display points upwards
1x
(aspiration).
e.
Aspirate liquid: Press the pipetting key once to
aspirate the liquid.
1x
f.
In order aspirate more liquid than set (up to max.
110% of the nominal volume), keep the pipetting
key pressed during the aspiration process until
the desired volume has been aspirated.
A rhombus appears in the display.
Hold pressed
'Mode' continues to flash.