6.7 Verification
General:
According to EU directive 2014/31/EU balances must be officially verified if they are
used as follows (legally controlled area):
• For commercial transactions if the price of goods is determined by weighing.
• For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
• For official purposes
• For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Balances in the legally controlled area (-> verified balances) must keep the error
limits in the verification validity period – normally they are the double of the
verification error limits.
When this verification validity period expires, a re-verification must be carried out.
Should be necessary an adjustment of the balance to keep the verification error limits
to satisfy the reverification requirements, this is not deemed a warranty case.
Verification notes:
An EU Qualification Approval is in existence for verified weighing systems. If a
balance is used where obligation to verify exists as described above, it must be
verified and re-verified at regular intervals.
Reverification is carried out according to the relevant national statutory regulations.
The validity for verification of balances in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Verification of the weighing system is invalid without the "seals".
The seal marks attached on balances with type approval point out that the
balance may only be opened and serviced by trained and authorised special-
ist staff. If the seal mark is destroyed, verification looses its validity. Please
observe all national laws and legal regulations. In Germany a re-verification
will be necessary.