Télécharger Imprimer la page

KERN KFS-TM Notice D'utilisation Et D'installation page 131

Publicité

Les langues disponibles

Les langues disponibles

8.2 Overview verifiable weighing systems
(in the configuration menu select the menu item F3 APP Setting „on")
Menu item
F0 SEL
Enable tolerance check
F1 Co
Display conditions of the tolerance
marker
F2 Li
Tolerance range
F3 Pn
Number of limiting points
F4 bU
Audio signal
F5 Ao
Automatic zero point correction
(zero tracking)
F6 AP
Automatic shutdown for battery
operation
F7 UA
RS-232 mode
KFS-TM-BA_IA-def-2321
Available settings
1 SEL0
Tolerance control disabled
1 SEL1
Tolerance control for weighing
1 SEL2*
Tolerance control for counting
Tolerance marker is always displayed, even if
11 Co0
standstill control is not yet displayed.
Tolerance marker is only displayed in connection
11 Co 1*
with standstill control.
Tolerance marker is only displayed above zero
12 Li 0
range.
Tolerance marker is displayed for the whole
12 Li 1*
range.
13 Pn 0
1- Limiting point (OK/ -)
13 Pn 1*
2- Limiting points (+/OK/-)
14 bu0*
Audio sound during tolerance control disabled
14 bu1
Audio sound when load is within tolerance limits
14 bu2
Audio sound when load is beyond tolerance limits
2 Ao0
Automatic zero tracking off
2 Ao1
Automatic limiting point correction on, 0.5 d
2 Ao2*
Automatic limiting point correction on, 1 d
2 Ao3
Automatic limiting point correction on, 2 d
2 Ao4
Automatic limiting point correction on, 4 d
3 Ap0*
AUTO OFF function disabled
Instrument will be switched off after 3 minutes of
3 Ap1
inactivity of display unit or weighing bridge.
4 UA0
Output via RS232C interface disabled
4 UA1*
Continuous data output
4 UA2
Continuous data output of stable weighing values
One output for stable weighing value. No output
4 UA3
for stable weighing values. Renewed output after
stabilization.
For remote commands,
4 UA4
Issue after pressing the PRINT ´button
Standard printer setting, output after pressing the
4 UA5
PRINT button
id on/off
dt on/off
G on/off
n on/off
C on/off
PCS on(off)
UW on/off
t on/off
4 UA6
Select TP-UP Printer or LP-50 Printer
4 UA7
KCP on/off
see chap. 9.2.
Printout memory on/off
Printout date on/off
Printout gross weight on/off
Printout net weight on/off
Printout total on/off
Printout parts counting on/off
Printout weighing unit on/off
Tara value printout
55

Publicité

loading