Call menu item linearization
Confirm menu setting
Ensure that there are no objects on the weighing plate.
"LoAd 1" will be displayed after standstill control has been carried out. Put the first
adjustment weight approx. 1/4 Max (see table 1) carefully in the centre of the
weighing pan.
"LoAd 2" will be displayed after standstill control has been carried out.
Put the second adjustment weight approx. 2/4 max (see table 1) carefully in the
centre of the weighing pan. "LoAd 3" will be displayed after standstill control has
been carried out.
Put the third adjustment weight approx. 3/4 max (see table 1) carefully in the
centre of the weighing pan. "LoAd 4" will be displayed after standstill control has
been carried out.
Put the forth adjustment weight approx. 4/4 max (see table 1) carefully in the
centre of the weighing pan.
After successful standstill control the balance carries out a selftest, then it
automatically returns to weighing mode.
•
An adjusting error or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure.