13.3 Automatic or manual transmission of the average item weight from
reference balance EWJ to bulk scales IFS
Make in the menu the following settings:
Switch on weighing scales and press and hold the MODE key during the self-
test until F1 Unt appears on the screen.
Press MODE key repeatedly until F3 Com in the display appears.
Confirm with 0 key, RS 232 will appear
Press again the 0-key, P Send will be shown
Mit (??) press 0-key, select Auto or mAnUAL* and acknowledge with 0-key
b 9600 will be displayed; acknowledge with 0-key y with PRINT/ESC return
into weighing mode
*
• mAnUAL: Transfer of the weight of a single piece to IFS scale after the
PRINT button is pressed.
• AUto: Weight of a single part is transferred to IFS scale automatically.
Define the average item weight:
Place the known item weight on the weighing plate of the EWJ
Press the PCS-key, the item number entered as last will be displayed,
e.g. SP 10.
Select the corresponding item number with MODE, e.g. SP 100, confirm with
the 0-key, ------- will be shortly displayed, followed by the set item number,
e.g. 200.
Transmit the average item weight to the bulk scales IFS:
Switch-on IFS with ON/OFF, press the F-key in weighing mode, the menu will
be invoked
Press the 8 key until SAmPLE is displayed
Confirm with the F-key, rS232 will be displayed
Press again the F-key, SAmPLE will be displayed again
Use +/- - key to return into the weighing mode
Place the weighing good on the platform of the IFS, the weight will be
displayed
The average item weight will be automatically transmitted to the IFS
The corresponding item number is automatically calculated and displayed.
KFS-TM-BA_IA-def-2321