Err. 46 – Plate safety temperature > plate temperature (more than + 40 K)
Causes
Effect
Solutions
If the actions described fails to resolve the fault or another error
code is displayed then take one of the following steps:
› contact the service department.
› send the device for repair, including a short description of the fault.
Warranty
/// Regulations
In accordance with IKA warranty conditions, the warranty period is 24
months. Upon registering this device at www.ika.com/register, your product
will be subject to a lifetime warranty. For claims under warranty, please cont-
act your local dealer.
The warranty does not cover worn out parts, nor does it apply to faults
resulting from improper use, insufficient care or maintenance not carried
out in accordance with the instructions in this operating manual.
› irregular temperature distribution across heating plate
due to sporadic heat dissipation
› defective control or safety temperature sensor
› heating switched off
› switch device off, leave to cool and switch on again
› ensure regular heat dissipation when using metal blocks,
etc. on the flat surface of the heating plate
Error codes | Warranty
55