/// Error 5
Providing safety protection, Error 5 is taken to recognize that the sensor
has not been placed in the medium when the heating is switched on.
Note! (
)
You can set a value from 1 to 30 min for this time limit depending on the
application. If time limit is set to 0 min, the Error 5 detection is disabled.
Caution! (
)
This function will only be active if:
› sensor temperature is < + 50 °C
› difference target temperature / sensor temperature > 5
/// Expiration of timer
You can specify the procedure of heating and stirring function if the timer
achieves 00:00:00. Next to the visually / audible information you can decide
independently if the heating and stirring should be stopped automatically if
the timer is expired or should continue.
Caution! (
)
If heating and stirring function is set to stop after timer is expired,
you have to ensure that there is no heat distortion for this application.
If heat distortion is not desirable, the stirring function should continue.