Operation:
Rinsing medium supply:
Preselect the rinsing medium supply
by pressing one of the left keys.
The selected tank is illuminated
from below.
Power setting in %:
Preselect the abrasive power in %
by pressing one of the keys in the
bottom row. The preselected pow-
er value is indicated digitally on the
display. You can increase the power
by:
a) pressing and holding the prese-
lected value or
b) pressing and holding a higher
value.
The power can be decreased by
pressing and holding a lower value.
Detection function:
Press key "1" shortly to switch on
and off.
Foot switch:
Boost function:
i.e. the preset power value is in-
creased by 20: Actuate the left ped-
al.
Normal operation
, i.e. abrasive
power and cooling/rinsing media:
both pedals must be pressed
simultaneously.
Key combinations:
Signal mode:
To switch the signal tone on or off, press "Small rinsing media tank" and
"PERIO 1" simultaneously.
The signal tone can be assigned to the detection of "calculus" with the
"PERIO 1" key or to "noncontact of the tip" with the root surface with the
"PERIO 2" key.
It is possible to adjust the volume of the signal with: "50%" (quieter) and
"100%" (louder).
Patient communication:
Press the "Large rinsing media tank" and "PERIO 2" keys simultaneously
to switch the function on or off.
To switch between images, press the "PERIO 1", "PERIO 2" and
"PERIO 3" keys.
Demonstration mode for sales reps:
To demonstrate the detection function with replacement materials.
Press "Fresh
water"
and "PERIO 3" simultaneously to switch on.
Press the foot switch to switch off.
Boost function:
Press keys "1" and "2" simultaneously to switch function on or off.
Reset:
Switch the unit off.
Then press and hold key "1" while switching the unit on.
All manuals and user guides at all-guides.com
PerioScan Quick Reference Guide
PerioScan
1%
25%
PERIO 1
PERIO 2
PERIO 3
Boost
50%
75%
100%
1
2
Reprogramming of program
keys:
Select e. g. the rinsing media sup-
ply type and power value in %.
Press the desired PERIO key for
more than 3 sec. until the indicator
bar next to the key flashes briefly.
The setting is safed.
Continual rinsing:
Press key "2" for 60 s continual
rinsing
Can be terminated by pressing the
foot switch or by pressing the key
again.
Learning program for
probing
pressure:
Press key "1" long to switch on and
off.
Can be terminated by pressing the
foot switch.
Rinsing:
when pressing the right pedal, only
rinsing medium is discharged from
the tip, flushing loose particles out of
the treatment area. The vibration of
the instrument is deactivated in this
case.
PerioScan
PERIO 1
PERIO 2
PERIO 3
Boost
1%
25%
50%
75%
100%
1
2