I – INTRODUCTION
Congratulations on the purchase of your new PIEZOTOME SOLO™ device.
To maximise use of this cutting-edge product for as long as possible, please read these instructions carefully
before switching on, using or servicing the device.
Sentences containing the symbol
Designed for clinical use:
Created by SATELEC, PIEZOTOME SOLO is a device designed to perform dental surgery including osteotomies,
osteoplasties, periodontal and implant surgery...
II - WARNINGS
CAUTION:
United States Federal Law restricts the use of this device solely to qualified, trained and competent dental
health practitioners or under their supervision.
The surgical tips designed for PIEZOTOME SOLO are not compatible with the PIEZOTOME and the IMPLANT
CENTER, and vice versa.
PIEZOTOME SOLO can only take tips adapted to PIEZOTOME 2 or IMPLANTCENTER 2.
To reduce the risk of accidents, the precautions stipulated below must be taken:
Device operators:
- Use of the device is restricted solely to graduate dental health professionals, who are competent and qualified
within the usual context of their activities.
- If you have received this device by mistake, please contact the device supplier to arrange for its return.
Interactions:
Interference may occur when used on patients with cardiac pacemakers.
This system emits electromagnetic fields, which means there are some potential risks. The
malfunctioning of implantable devices such as cardiac pacemakers and ICDs (implantable cardioverter
defibrillator) is possible:
- Ask patients and users if they have an implanted device before using this product. Explain the circumstances to
them.
- Weigh the risks and benefits and contact your patient's cardiologist or appropriate qualified healthcare
professional prior to performing the treatment.
- Keep this product away from implanted devices.
- Make appropriate emergency provisions and take immediate action if patients become ill.
- Symptom including a raised heartbeat, irregular pulse, and dizziness may signal problems with a cardiac
pacemaker or ICD.
Electrical connection:
- Ask an approved dental installation technician to connect your device to the electrical network.
- Warning: To prevent any risk of electric shock, this device shall only be connected to a supply network fitted
with a protective ground.
- The electrical network to which the device is connected shall comply with the standards in force in your
country.
- If loss of electrical power during use is likely to generate an unacceptable risk, the device must be connected
to a suitable power source (e.g. UPS).
- Do not attempt to connect to the device any other accessories than those supplied by SATELEC.
Using the device:
- Do not use the device if it appears to be damaged or faulty.
- Before unplugging the mains power cable, switch the device off.
- To unplug the mains power cable, pull the plug on the end of the cable whilst holding the wall socket with your
other hand.
- Never use irrigating solution containers that are not designed to be suspended from the bracket provided.
- The device must only be used with bottles or pouches of sterile water or physiological saline solution.
- The capacity of the irrigating solution container used shall not exceed one litre.
- If you are not using or will be away from the device for a period of time, disconnect the device from the
electrical supply network.
- Do not move the device during use.
Piezotome Solo V6 V2
are points of particular importance.