Note:
Automatic storage allows you to save the display parameters used previously when switching between the
different programmes.
This sequence is carried under controlled aseptic conditions for the safety of your patients. You may require the
help of an assistant.
- Connect the handpiece cable to the device connector (Fig. 1, item 7).
- Screw the selected tip onto the handpiece using a wrench (according to the recommendations of the clinical
manual).
- Attach the irrigation line to the cable using line clips.
- Attach the end of the irrigation line to the handpiece.
- Prime the irrigation system by pressing the flush button on the screen (Fig. 3, item 3)
- When the irrigation solution arrives at the tip of the tip, release the pedal.
- Use the device in compliance with good dentistry practice.
X – SWITCHING OFF THE DEVICE
At the end of the dental procedure:
- Remove the bottle or pouch containing physiological saline solution or sterile water from the bracket.
- Remove the irrigation line perforator from the bottle or pouch.
- Immerse the irrigation line perforator in a container of distilled water.
- Rinse the irrigation line and the handpiece by activating the flush function until the receptacle and the
irrigation line are completely empty.
- Set the device to OFF (O).
- Disconnect the handpiece from the device.
- Remove the irrigation line clips.
- Disconnect the irrigation line from the handpiece and throw it in a bin used for the disposal of used medical
items (irrigation line can only be used once).
XI – SERVICING AND STERILIZATION

11.1 SERVICING THE DEVICE

Important:
The device must be OFF during cleaning/disinfection procedures.
Do not use abrasive product, sprays or liquids to clean and disinfect the console.
Avoid the use of cleaning and disinfection products containing flammable agents.
If you do use such a product, make sure that all the product has evaporated and that no flammable agents
remain on the device and its accessories before using the device again.
The device, the bracket and the multifunction pedal are not suitable for sterilization.
The console, the control pedal and the bracket shall be cleaned and disinfected (using alcohol, disinfection
products, dentist disinfecting wipes, SEPTOL™ wipes) systematically after each use.
The console must be checked regularly to detect any problems.
It is important that you check the cleanliness of the ventilation inlets on the console to prevent abnormal
heating.

11.2 SERVICING THE ULTRASOUND HANDPIECE

Important:
The handpiece is not delivered sterile.
Refer to the user's manual delivered with the equipment.

11.3 SERVICING TIPS

Important:
Tips are not delivered sterile.
Refer to the user's manual delivered with the equipment.
Piezotome Solo V6 V2
11