• Remove any externally attached soiling by brushing carefully with a soft surface brush or cloth.
• Soak the Easy Clean in the cleaning solution for, as a minimum, the time and the concentration specifi ed by the manufacturer of the cleaning agent (ENZOL,
2 %, 10 min).
• Brush the lumen and surface with a suitable lumen / surface brush until all visible soiling is removed.
• Rinse through the lumina with the cleaning solution (ENZOL, 2 %), at least 5 times using a disposable syringe (min. volume 50 ml).
• Submerge the Easy Clean completely in deionized water.
• Rinse through the lumina with deionized water at least 3 times using a disposable syringe (min. volume 50 ml).
• Repeat the cleaning process if the last rinse does not run clear, or if stains are still visible on the product.
• Dry the Easy Clean at room temperature.
• Place the Easy Clean in a sterilization bag.
The Easy Clean can be cleaned and sterilized in a fractional pre vacuum cycle at 132°C for 4 minutes.