3. Operation
Common operation
1. Place the open end of the tube (or use the supplied mouth piece) in the mouth or close to
opening of the tracheotomy tube
2. Select the aerosol output as required Push +/- to in/decrease function in steps
3. Set air flow as required.Push +/- to in/decrease function in steps
Caution - Take care of enough liquid inside nebulizer chamber and sterile water system.
Caution - Do not allow the nebuliser chamber to run dry !
If the unit runs dry, it will automatically switch off automatically, the red LED will light and
audio alarm will sounds. After 2 min. the audio alarm will fall silent.
Caution - Red LED and audio alarm indicates a malfunction.
Press the on/off switch: Audio Alarm and Red LED goes out
The green LED is lit, refill chamber if needed
Press O/I to continue if needed
If a fault continuous check troubleshooting Guide
If the Is a malfunction not rectified call DeVilbiss or your local authorised dealer
Timer setting
1. Push the time button to increase the timer in 15 min increments. The maximum time is
60 minutes.
2. Push +/- to in/decrease function in steps (steps are 15-30-45-60-0 min.)
3. If using sterile water system without continuous filling set, don't exceed the max. time of
15 minutes.
4. If the timer is active the yellow LED will light. Once the set time has been exceeded the
UltraNeb will switch off automatically
5. Red light and audio warning will come on for 2 min.
6. To restart the device, push the on/off switch O/I
Shut-down
1. Push the O/I button to set the machine to stand-by mode This switches off transducer,
fan and hose heating.
2. Green LED light diode is lit.
3. close the knurled knob on the plastic mechanise on the filling tube
4. For long turn shut down set the unit from standby to power down by pressing the O/I
switch on the rear panel to the O mode
English