Warning on power connections
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each country
if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power Cord /
Appliance Connector / Plug, please consult a qualified
service personnel.
Warning on power connections for
medical use
Customers in the U.S.A. and Canada should use the
following type of power cord.
Customers in other countries or regions should use the
power cord prescribed by their country or region.
Plug type
Cord type
Minimum rating for plug
and appliance couplers
Safety approval
* Note: Grounding reliability can only be achieved when
the equipment is connected to an equivalent receptacle
marked "Hospital Only" or "Hospital Grade".
Caution
• This unit is heavy. Make sure to unpack and move the
unit with two or more people.
• Firmly grip the bottom of this unit as shown below.
10
GB
U.S.A. and Canada
*
HOSPITAL GRADE
Min. Type SJT
Min. 18 AWG
10 A / 125 V
UL Listed and CSA
Caution
Do not come into contact with the terminals of the rear
panel connectors and patients at the same time.
Doing so may result in a generation of voltage that can be
harmful to patients if the unit is malfunctioning.
Always disconnect the power cord before connecting and
disconnecting connectors.
For details on operation, refer to the manuals
stored on the supplied CD-ROM.
Using the CD-ROM manuals
The manuals can be viewed on a computer with Adobe
Reader installed.
You can download Adobe Reader for free from the
Adobe website.
1. Open the "index.html" file on the CD-ROM.
2. Select the language of the manual you want to view.
Note
If you damage or lose the CD-ROM, you can purchase a
new one from your dealer or Sony service
representative.