Important EMC notices for use in medical environments
• The LMD-X310S/X550S needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in the instructions for use.
• The LMD-X310S/X550S is intended for use in a professional healthcare facility environment.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the LMD-X310S/X550S.
Warning
• Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-X310S/
X550S. Otherwise, degradation of the performance of this equipment could result.
• If the LMD-X310S/X550S will be used adjacent to or stacked with other equipment, normal operation of the LMD-
X310S/X550S under such configurations should be verified via observation.
• The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased immunity of the LMD-X310S/X550S.
Guidance and manufacturer's declaration – electromagnetic emissions
The LMD-X310S/X550S is intended for use in the electromagnetic environment specified below. The customer or the
user of the LMD-X310S/X550S should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
4
GB
Compliance
The LMD-X310S/X550S uses RF energy only for its
internal function. Therefore, its RF emissions are very low
Group 1
and are not likely to cause any interference in nearby
electronic equipment.
The LMD-X310S/X550S is suitable for use in all
Class B
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
Not applicable
purposes.
(LMD-X310S)
Class D
(LMD-X550S)
Not applicable
(LMD-X310S)
Complies
(LMD-X550S)
Electromagnetic environment – guidance