4. Acquiring CEPH Images (Optional)
4.2.5
Carpus Mode
For Carpus Mode, installing the carpus plate is required. Follow the procedure below
to position the patient for the mode.
1.
Turn the Nasal Positioner to the PA/Waters' view / Carpus mode marker, as
shown below.
2.
Fold the Nasal Positioner up. The Nasal Positioner is not used in Carpus mode.
3.
Slide the two ends of the carpus plate into the two holes on the CEPH unit as
below. Ensure to check that the carpus plate is fully locked in the position before
moving on to the next step.
4.
Ask the patient to place his or her right-hand flat on the carpus plate. All fingers
must be stretched out and remain in the position until the scanning is complete.
5.
Click the READY button. The X-ray will not be emitted at this point.
6.
Go to 4.3 X-ray Exposure and follow the procedure for the X-ray exposure.
52
TM
Green X 18
(Model: PHT-75CHS) User Manual