2
Making Measurements
Measuring Resistance, Conductance, and Testing Continuity
Disconnect circuit power and discharge all high-voltage capacitors before
CAUTION
measuring resistance or conductance, or testing circuit continuity, to avoid
damaging the multimeter or the device under test.
Set up the multimeter to measure resistance as shown in
the test points (by shunting the resistor) and read the display.
Press
NOTE
Setup configuration), conductance measurement (
measurement (
Smart
Using the offset compensation method, Smart removes unexpected DC
voltages within instrument, at the input, or the circuit being measured, which will
add error to resistance measurement. Besides, it also displays the bias voltage or
leakage current (calculated based on bias voltage and corrected resistance value)
on the secondary display. With offset compensation method, the multimeter takes
the difference between two resistance measurements when two different test
currents are applied to determine any offset voltage in the input circuitry. The
resultant displayed measurement corrects this offset, giving a more accurate
resistance measurement.
64
to scroll through audible continuity test (
,
, or
) as shown in
Figure 2-9
Keysight U1253B User's and Service Guide
Figure
2-8. Then probe
or
, depending on
), and resistance
on page 67.